Transcutaneous blood gas monitoring in neonates
Transcutaneous blood gas monitoring uses electrodes on the skin to estimate the oxygen and carbon dioxide in the blood without taking a blood sample. A 2005 survey article says skin readings allow continuous monitoring, while conventional blood gas tests cause significant blood loss in preterm babies. Its survey of 41 neonatal units in German-speaking countries found the method widely used, but only 8% of units checked the skin readings against arterial blood gases. A 2016 study of 248 babies born before 33 weeks found a poor match between the skin carbon dioxide reading and the blood value, with the trend agreeing only 39.6% of the time. This page does not cover capillary blood gas sampling and gives no target values.
Also called Transcutaneous PCO2 and PO2 monitoring · tcPCO2 and tcPO2 monitoring · Noninvasive blood gas monitoring in newborns
- Technology
- In this atlas
Overview
Technical summary
Technology page (MeSH D001785, Blood Gas Monitoring, Transcutaneous: exact). Sources read: the MeSH preferred concept scope note (M0002691), the abstract of a 2005 survey of 41 German-speaking neonatal units (PMC1192805, BMC Pediatrics, CC BY 2.0) and the abstract of a 2016 retrospective single-unit study of transcutaneous carbon dioxide in 248 infants born before 33 weeks (PMC4916268, Critical Care Research and Practice, CC BY 4.0). The plan named transcutaneous or capillary blood gas; no source read here covers capillary blood gas sampling in neonates, and the MedlinePlus arterial blood gas page is about arterial sampling and not infants, so capillary sampling is left out. The 2005 survey predates current devices. No target values, sensor settings or alarm limits are stated. Recipient: newborn (preterm).
How it works
MeSH defines transcutaneous blood gas monitoring as the noninvasive measurement of the partial pressure of oxygen and carbon dioxide locally in the capillaries of a tissue, by a set of electrodes applied to the skin.
Strengths
- A 2005 survey article says transcutaneous measurements allow continuous, noninvasive monitoring of blood gas levels, whereas conventional blood gas measurements cause significant blood loss in preterm infants.
Limitations
- In a 2016 retrospective study of 248 infants born before 33 weeks, the Pearson correlation between transcutaneous and blood carbon dioxide values was 0.58 and trend agreement was good in only 39.6%; the authors warn against clinical decisions on the transcutaneous value alone.
- In the 2005 survey of 41 German-speaking units, only 8% obtained arterial blood gases to validate transcutaneous values, and target oxygen levels varied widely between units.
Required infrastructure
- A neonatal intensive care unit with transcutaneous monitors; the 2005 survey and the 2016 study were both set in neonatal intensive care units.
Claims and evidence
Context A vocabulary definition (MeSH), not a study population
- Recipient
- No clinical recipient
- Population
- Not applicable: a controlled-vocabulary definition from the National Library of Medicine, with no population.
- Care setting
- Not applicable (vocabulary)
- Geography
- Not applicable (vocabulary)
- Life stage
- Not applicable (vocabulary)
- Follow-up
- Not applicable (vocabulary)
Eligibility: 1 inclusion criterion; exclusion criteria not stated
- Included
- Not applicable (vocabulary)
- Excluded
- None stated
The National Library of Medicine vocabulary MeSH defines transcutaneous blood gas monitoring as the noninvasive measurement of the partial pressure of oxygen and/or carbon dioxide locally in the capillaries of a tissue, by applying a special set of electrodes to the skin.
Applies to No clinical recipient A vocabulary definition (MeSH), not a study population
- Certainty method
- Not assessed
- Limitations
- A vocabulary definition, not a clinical criterion; it covers all ages and does not mention newborns. MeSH wording describes the electrodes' photoelectric sensors in more detail, which this page does not repeat.
- Last checked
- 2026-11-18
- Next evidence check due
- Not scheduled
Sources and citations (1)
-
- Locator
- scopeNote.@value
- How it applies
- Passage in the source that backs the claim.
- Retrieved
- Source effective date
- Not reported
- Source state
- Current
- Version fingerprint
09d7f7f69b1530316acd2d80ee5eabc7806a21220f22b6a36dc32b5fb2130889- Reuse basis
- The NLM web policies page, fetched 2026-11-08 (sha256 c8b7026b5c3fcfd1b0cb88755077f6a99a5f2159eecd42b0e7d1a5ad9b30ec55, hash only), says works produced by the U.S. government are not subject to copyright protection in the United States and asks users to acknowledge NLM as the source. Source: National Library of Medicine. The NLM terms and conditions page, fetched 2026-11-08 (sha256 f1fb494447a54d2190e496410163c8e1fc138a601d83d8c124f267f7103026b3, hash only), asks redistributors to maintain the most current data or to make known that their copy may not reflect the most current data. This copy was fetched on 2026-11-18 and may not reflect the most current data. The raw bytes are committed in the atlas repository.
Context Neonatal intensive care units in Germany, Switzerland and Austria surveyed about transcutaneous blood gas monitoring (41 units answered)
- Recipient
- Newborn
- Population
- Neonatal intensive care units: a questionnaire went to 56 head nurses and 41 units gave a complete answer. The survey asks about unit practice and does not describe patients.
- Care setting
- Neonatal intensive care units
- Geography
- Germany, Switzerland and Austria
- Life stage
- Newborn intensive care, with a statement about preterm infants
- Gestational age
- Not reported
- Postnatal age
- Not reported
- Follow-up
- Not applicable (a survey of unit practice)
Eligibility: 1 inclusion criterion; exclusion criteria not stated
- Included
- German-speaking neonatal intensive care units sent the questionnaire
- Excluded
- None stated
A 2005 survey article says that transcutaneous measurements allow continuous, noninvasive monitoring of blood gas levels, while conventional blood gas measurements cause significant blood loss in preterm infants. A questionnaire sent to 56 head nurses in German-speaking neonatal intensive care units was answered completely by 41; transcutaneous measurements were not performed in two of them, and in most units (77%) oxygen and carbon dioxide were measured simultaneously. Only 8% of units obtained arterial blood gases to validate the transcutaneous values. The authors conclude that use of transcutaneous monitors remains widespread in these units, despite earlier data suggesting it had been abandoned in many units worldwide, and that the current way of monitoring oxygenation may not prevent hyperoxaemia in preterm infants.
Applies to Newborn Neonatal intensive care units in Germany, Switzerland and Austria surveyed about transcutaneous blood gas monitoring (41 units answered)
- Certainty method
- Not assessed
- Limitations
- A 2005 questionnaire from three countries, so it reports practice then, not now, and nurses' answers rather than measured outcomes. It gives no accuracy data. It reports wide variation between units in target oxygen levels, which this page does not repeat as values. Recipient: newborn (neonatal unit practice).
- Last checked
- 2026-11-18
- Next evidence check due
- Not scheduled
Sources and citations (7)
-
Supports A survey of transcutaneous blood gas monitoring among European neonatal intensive care units
- Locator
- Abstract, Background, sentence 2
- How it applies
- Passage in the source that backs the claim.
- Retrieved
- Source effective date
- Not reported
- Source state
- Current
- Version fingerprint
73647a5c0905e8f02bd37a2619b3f08fd1da21bacbc911baa6bf799550f06a13- Reuse basis
- PMC Open Access subset article under https://creativecommons.org/licenses/by/2.0/ (CC BY 2.0; the article's own licence element in the XML). The raw XML from NCBI E-utilities is committed unaltered in the atlas repository, which CC BY 2.0 allows with attribution. Short quotations under 25 words.
-
Supports A survey of transcutaneous blood gas monitoring among European neonatal intensive care units
- Locator
- Abstract, Methods
- How it applies
- Passage in the source that backs the claim.
- Retrieved
- Source effective date
- Not reported
- Source state
- Current
- Version fingerprint
73647a5c0905e8f02bd37a2619b3f08fd1da21bacbc911baa6bf799550f06a13- Reuse basis
- PMC Open Access subset article under https://creativecommons.org/licenses/by/2.0/ (CC BY 2.0; the article's own licence element in the XML). The raw XML from NCBI E-utilities is committed unaltered in the atlas repository, which CC BY 2.0 allows with attribution. Short quotations under 25 words.
-
Supports A survey of transcutaneous blood gas monitoring among European neonatal intensive care units
- Locator
- Abstract, Results, sentences 1 and 2
- How it applies
- Passage in the source that backs the claim.
- Retrieved
- Source effective date
- Not reported
- Source state
- Current
- Version fingerprint
73647a5c0905e8f02bd37a2619b3f08fd1da21bacbc911baa6bf799550f06a13- Reuse basis
- PMC Open Access subset article under https://creativecommons.org/licenses/by/2.0/ (CC BY 2.0; the article's own licence element in the XML). The raw XML from NCBI E-utilities is committed unaltered in the atlas repository, which CC BY 2.0 allows with attribution. Short quotations under 25 words.
-
Supports A survey of transcutaneous blood gas monitoring among European neonatal intensive care units
- Locator
- Abstract, Results, sentence 3
- How it applies
- Passage in the source that backs the claim.
- Retrieved
- Source effective date
- Not reported
- Source state
- Current
- Version fingerprint
73647a5c0905e8f02bd37a2619b3f08fd1da21bacbc911baa6bf799550f06a13- Reuse basis
- PMC Open Access subset article under https://creativecommons.org/licenses/by/2.0/ (CC BY 2.0; the article's own licence element in the XML). The raw XML from NCBI E-utilities is committed unaltered in the atlas repository, which CC BY 2.0 allows with attribution. Short quotations under 25 words.
-
Supports A survey of transcutaneous blood gas monitoring among European neonatal intensive care units
- Locator
- Abstract, Results, sentence 5
- How it applies
- Passage in the source that backs the claim.
- Retrieved
- Source effective date
- Not reported
- Source state
- Current
- Version fingerprint
73647a5c0905e8f02bd37a2619b3f08fd1da21bacbc911baa6bf799550f06a13- Reuse basis
- PMC Open Access subset article under https://creativecommons.org/licenses/by/2.0/ (CC BY 2.0; the article's own licence element in the XML). The raw XML from NCBI E-utilities is committed unaltered in the atlas repository, which CC BY 2.0 allows with attribution. Short quotations under 25 words.
-
Supports A survey of transcutaneous blood gas monitoring among European neonatal intensive care units
- Locator
- Abstract, Conclusion, sentence 1
- How it applies
- Passage in the source that backs the claim.
- Retrieved
- Source effective date
- Not reported
- Source state
- Current
- Version fingerprint
73647a5c0905e8f02bd37a2619b3f08fd1da21bacbc911baa6bf799550f06a13- Reuse basis
- PMC Open Access subset article under https://creativecommons.org/licenses/by/2.0/ (CC BY 2.0; the article's own licence element in the XML). The raw XML from NCBI E-utilities is committed unaltered in the atlas repository, which CC BY 2.0 allows with attribution. Short quotations under 25 words.
-
Supports A survey of transcutaneous blood gas monitoring among European neonatal intensive care units
- Locator
- Abstract, Conclusion, sentence 2
- How it applies
- Passage in the source that backs the claim.
- Retrieved
- Source effective date
- Not reported
- Source state
- Current
- Version fingerprint
73647a5c0905e8f02bd37a2619b3f08fd1da21bacbc911baa6bf799550f06a13- Reuse basis
- PMC Open Access subset article under https://creativecommons.org/licenses/by/2.0/ (CC BY 2.0; the article's own licence element in the XML). The raw XML from NCBI E-utilities is committed unaltered in the atlas repository, which CC BY 2.0 allows with attribution. Short quotations under 25 words.
Context Premature babies born before 33 weeks in one tertiary neonatal intensive care unit (248 babies, 1365 paired measurements)
- Recipient
- Newborn
- Population
- 248 consecutive premature infants born before 33 weeks gestational age (median gestational age 29 weeks 5 days, median birth weight 1250 g), who gave 1365 pairs of transcutaneous and blood carbon dioxide values.
- Care setting
- A tertiary neonatal intensive care unit
- Geography
- Not stated in the abstract
- Life stage
- Premature newborn period; the abstract gives no postnatal ages for the measurements
- Gestational age
- Not reported
- Postnatal age
- Not reported
- Follow-up
- Paired transcutaneous and blood measurements during intensive care; a retrospective study
Eligibility: 1 inclusion criterion; exclusion criteria not stated
- Included
- All consecutive premature infants born before 33 weeks gestational age in the unit
- Excluded
- None stated
A 2016 retrospective study in one tertiary neonatal intensive care unit analysed 1365 pairs of transcutaneous carbon dioxide (TcpCO2) and blood carbon dioxide (pCO2) values from 248 consecutive premature infants born before 33 weeks (median gestational age 29 weeks 5 days, median birth weight 1250 g). The Pearson correlation between the values was 0.58, the mean bias was -0.93 kPa with 95% limits of agreement of -4.05 to +2.16 kPa, and the trends of the two values agreed well in only 39.6%. The authors conclude that in premature infants TcpCO2 was poorly correlated with blood pCO2, and warn about clinical decisions based on TcpCO2 alone when it is used as continuous monitoring.
Applies to Newborn Premature babies born before 33 weeks in one tertiary neonatal intensive care unit (248 babies, 1365 paired measurements)
- Certainty method
- Not assessed
- Limitations
- Retrospective and from a single unit, with one device type that the abstract does not name; the results may not apply to other devices, term infants or oxygen readings. The abstract does not give postnatal ages for the measurements. Recipient: newborn (preterm, born before 33 weeks).
- Last checked
- 2026-11-18
- Next evidence check due
- Not scheduled
Sources and citations (8)
-
Supports Accuracy of Transcutaneous Carbon Dioxide Measurement in Premature Infants
- Locator
- Abstract, Background, sentence 2
- How it applies
- Passage in the source that backs the claim.
- Retrieved
- Source effective date
- Not reported
- Source state
- Current
- Version fingerprint
dee91e39567bf3453903108cd83d7718027dda57db2fd40e974f3e01167168af- Reuse basis
- PMC Open Access subset article under https://creativecommons.org/licenses/by/4.0/ (CC BY 4.0; the article's own licence element in the XML). The raw XML from NCBI E-utilities is committed unaltered in the atlas repository, which CC BY 4.0 allows with attribution. Short quotations under 25 words.
-
Supports Accuracy of Transcutaneous Carbon Dioxide Measurement in Premature Infants
- Locator
- Abstract, Methods, sentence 1
- How it applies
- Passage in the source that backs the claim.
- Retrieved
- Source effective date
- Not reported
- Source state
- Current
- Version fingerprint
dee91e39567bf3453903108cd83d7718027dda57db2fd40e974f3e01167168af- Reuse basis
- PMC Open Access subset article under https://creativecommons.org/licenses/by/4.0/ (CC BY 4.0; the article's own licence element in the XML). The raw XML from NCBI E-utilities is committed unaltered in the atlas repository, which CC BY 4.0 allows with attribution. Short quotations under 25 words.
-
Supports Accuracy of Transcutaneous Carbon Dioxide Measurement in Premature Infants
- Locator
- Abstract, Methods, sentence 2
- How it applies
- Passage in the source that backs the claim.
- Retrieved
- Source effective date
- Not reported
- Source state
- Current
- Version fingerprint
dee91e39567bf3453903108cd83d7718027dda57db2fd40e974f3e01167168af- Reuse basis
- PMC Open Access subset article under https://creativecommons.org/licenses/by/4.0/ (CC BY 4.0; the article's own licence element in the XML). The raw XML from NCBI E-utilities is committed unaltered in the atlas repository, which CC BY 4.0 allows with attribution. Short quotations under 25 words.
-
Supports Accuracy of Transcutaneous Carbon Dioxide Measurement in Premature Infants
- Locator
- Abstract, Results, sentence 1
- How it applies
- Passage in the source that backs the claim.
- Retrieved
- Source effective date
- Not reported
- Source state
- Current
- Version fingerprint
dee91e39567bf3453903108cd83d7718027dda57db2fd40e974f3e01167168af- Reuse basis
- PMC Open Access subset article under https://creativecommons.org/licenses/by/4.0/ (CC BY 4.0; the article's own licence element in the XML). The raw XML from NCBI E-utilities is committed unaltered in the atlas repository, which CC BY 4.0 allows with attribution. Short quotations under 25 words.
-
Supports Accuracy of Transcutaneous Carbon Dioxide Measurement in Premature Infants
- Locator
- Abstract, Results, sentence 2
- How it applies
- Passage in the source that backs the claim.
- Retrieved
- Source effective date
- Not reported
- Source state
- Current
- Version fingerprint
dee91e39567bf3453903108cd83d7718027dda57db2fd40e974f3e01167168af- Reuse basis
- PMC Open Access subset article under https://creativecommons.org/licenses/by/4.0/ (CC BY 4.0; the article's own licence element in the XML). The raw XML from NCBI E-utilities is committed unaltered in the atlas repository, which CC BY 4.0 allows with attribution. Short quotations under 25 words.
-
Supports Accuracy of Transcutaneous Carbon Dioxide Measurement in Premature Infants
- Locator
- Abstract, Results, sentence 3
- How it applies
- Passage in the source that backs the claim.
- Retrieved
- Source effective date
- Not reported
- Source state
- Current
- Version fingerprint
dee91e39567bf3453903108cd83d7718027dda57db2fd40e974f3e01167168af- Reuse basis
- PMC Open Access subset article under https://creativecommons.org/licenses/by/4.0/ (CC BY 4.0; the article's own licence element in the XML). The raw XML from NCBI E-utilities is committed unaltered in the atlas repository, which CC BY 4.0 allows with attribution. Short quotations under 25 words.
-
Supports Accuracy of Transcutaneous Carbon Dioxide Measurement in Premature Infants
- Locator
- Abstract, Results, sentence 4
- How it applies
- Passage in the source that backs the claim.
- Retrieved
- Source effective date
- Not reported
- Source state
- Current
- Version fingerprint
dee91e39567bf3453903108cd83d7718027dda57db2fd40e974f3e01167168af- Reuse basis
- PMC Open Access subset article under https://creativecommons.org/licenses/by/4.0/ (CC BY 4.0; the article's own licence element in the XML). The raw XML from NCBI E-utilities is committed unaltered in the atlas repository, which CC BY 4.0 allows with attribution. Short quotations under 25 words.
-
Supports Accuracy of Transcutaneous Carbon Dioxide Measurement in Premature Infants
- Locator
- Abstract, Conclusions, sentences 1 and 3
- How it applies
- Passage in the source that backs the claim.
- Retrieved
- Source effective date
- Not reported
- Source state
- Current
- Version fingerprint
dee91e39567bf3453903108cd83d7718027dda57db2fd40e974f3e01167168af- Reuse basis
- PMC Open Access subset article under https://creativecommons.org/licenses/by/4.0/ (CC BY 4.0; the article's own licence element in the XML). The raw XML from NCBI E-utilities is committed unaltered in the atlas repository, which CC BY 4.0 allows with attribution. Short quotations under 25 words.