Pulse oximetry for newborns
A pulse oximeter is a small clip-like device that uses light through the skin to estimate how much oxygen the blood carries. In newborns it is used as a screening test for critical congenital heart defects before the baby leaves hospital. A 2021 review found consistent accuracy for detecting these defects when screening was added to scans in pregnancy and the clinical examination, with false positives judged not serious. A 2015 Malaysian study of 5247 well term newborns found that most babies with a positive screen had other illnesses that needed treatment, mainly breathing problems and sepsis, and that the test missed more than half of such illnesses. This page gives no oxygen thresholds.
Also called Neonatal pulse oximetry · Newborn oxygen saturation monitoring · Pulse ox
- Technology
- In this atlas
Overview
Technical summary
Technology page for the device and its newborn use. Sources read: the MedlinePlus pulse oximetry medical test page (general, no newborn content, used for the device description only), the abstract and one body paragraph of a 2021 review of newborn pulse oximetry screening recommendations and systematic reviews (PMC8424789, BMC Pediatrics, CC BY 4.0) and the abstract of a 2015 pilot study of pulse oximetry screening for non-cardiac hypoxaemic disease at one Malaysian hospital (PMC4567069, PLoS One, CC BY 4.0). The related record maternal:pulse-oximetry-cchd-screening states the MedlinePlus screening claim; this page does not repeat it. In the Malaysian study the 100% sensitivity for critical congenital heart disease rests on two cases. MeSH D010092 (Oximetry) is a broad identifier. No saturation thresholds are stated. Recipient: newborn.
How it works
A clip-like sensor on the skin uses light to estimate the oxygen level of the blood (MedlinePlus, general page). In newborn screening a pulse oximeter can read the saturation before the arterial duct (probe on the right hand) and after it (probe on either foot), as a 2021 review describes.
Classification codes
-
D010092Oximetry
Strengths
- A 2021 review found consistent accuracy for detecting critical congenital heart defects in newborns when pulse oximetry screening is added to antenatal ultrasonography and clinical examination, and judged the potential harms of false-positive tests not serious.
- MedlinePlus describes the test as painless and quick and says there is no known risk.
Limitations
- In a 2015 Malaysian pilot study the sensitivity for non-cardiac disease was 42%: 18 babies with negative screens were later admitted for sepsis or pneumonia.
- MedlinePlus lists factors that can make a result inaccurate in people in general, including darker skin pigmentation, poor circulation, cold skin and dark or artificial nails; the page does not mention newborns.
- The 100% sensitivity for critical congenital heart disease in the Malaysian study rests on two babies.
Required infrastructure
- A pulse oximeter and a trained tester. In the Malaysian study, babies with two positive screens an hour apart had echocardiography, so access to echocardiography follows a positive screen there.
Claims and evidence
Context Apparently healthy term newborns screened by pulse oximetry before discharge (5247 babies at one hospital in Malaysia)
- Recipient
- Newborn
- Population
- 5247 apparently healthy term newborns delivered at one university medical centre, screened before discharge. The median age at screening was 20 hours.
- Care setting
- A university hospital maternity service
- Geography
- Malaysia (a middle-income country)
- Life stage
- Term newborns screened at a median age of 20 hours, before discharge
- Gestational age
- Not reported
- Postnatal age
- Not reported
- Follow-up
- Echocardiography for babies with positive screening results on two occasions one hour apart; babies with negative screens were later admitted for illness and counted
Eligibility: 1 inclusion criterion; exclusion criteria not stated
- Included
- Apparently healthy term newborns delivered at the centre
- Excluded
- None stated
In a pilot study at one university hospital in Malaysia, 5247 apparently healthy term newborns were screened with pulse oximetry before discharge, at a median age of 20 hours, and 15 had positive results. Thirteen babies (0.24%) had significant non-cardiac disease that needed hospital treatment (2 with sepsis and 11 with respiratory disease) and two had critical congenital heart disease. A further 18 babies with negative screens were later admitted for sepsis (16) or pneumonia (2). The authors report a sensitivity of 42% and a specificity of 99.9% for non-cardiac disease, and 100% and 99.7% for critical congenital heart disease.
Applies to Newborn Apparently healthy term newborns screened by pulse oximetry before discharge (5247 babies at one hospital in Malaysia)
- Certainty method
- Not assessed
- Limitations
- A single-centre pilot in a middle-income country. The critical congenital heart disease figures rest on two babies. Only apparently healthy term newborns were screened. Positive meant positive results on two separate screens one hour apart, which is a protocol feature, not a general rule. Recipient: newborn (term).
- Last checked
- 2026-11-18
- Next evidence check due
- Not scheduled
Sources and citations (8)
-
- Locator
- Abstract, Methods and Findings, sentence 1
- How it applies
- Passage in the source that backs the claim.
- Retrieved
- Source effective date
- Not reported
- Source state
- Current
- Version fingerprint
0ace401db5d4f38c63e0108aab2fcbe976f2ad65e7e580d02f2def2833f917a0- Reuse basis
- PMC Open Access subset article under https://creativecommons.org/licenses/by/4.0/ (CC BY 4.0; the article's own licence element in the XML). The raw XML from NCBI E-utilities is committed unaltered in the atlas repository, which CC BY 4.0 allows with attribution. Short quotations under 25 words.
-
- Locator
- Abstract, Methods and Findings, sentence 2
- How it applies
- Passage in the source that backs the claim.
- Retrieved
- Source effective date
- Not reported
- Source state
- Current
- Version fingerprint
0ace401db5d4f38c63e0108aab2fcbe976f2ad65e7e580d02f2def2833f917a0- Reuse basis
- PMC Open Access subset article under https://creativecommons.org/licenses/by/4.0/ (CC BY 4.0; the article's own licence element in the XML). The raw XML from NCBI E-utilities is committed unaltered in the atlas repository, which CC BY 4.0 allows with attribution. Short quotations under 25 words.
-
- Locator
- Abstract, Methods and Findings, sentence 4
- How it applies
- Passage in the source that backs the claim.
- Retrieved
- Source effective date
- Not reported
- Source state
- Current
- Version fingerprint
0ace401db5d4f38c63e0108aab2fcbe976f2ad65e7e580d02f2def2833f917a0- Reuse basis
- PMC Open Access subset article under https://creativecommons.org/licenses/by/4.0/ (CC BY 4.0; the article's own licence element in the XML). The raw XML from NCBI E-utilities is committed unaltered in the atlas repository, which CC BY 4.0 allows with attribution. Short quotations under 25 words.
-
- Locator
- Abstract, Methods and Findings, sentence 5
- How it applies
- Passage in the source that backs the claim.
- Retrieved
- Source effective date
- Not reported
- Source state
- Current
- Version fingerprint
0ace401db5d4f38c63e0108aab2fcbe976f2ad65e7e580d02f2def2833f917a0- Reuse basis
- PMC Open Access subset article under https://creativecommons.org/licenses/by/4.0/ (CC BY 4.0; the article's own licence element in the XML). The raw XML from NCBI E-utilities is committed unaltered in the atlas repository, which CC BY 4.0 allows with attribution. Short quotations under 25 words.
-
- Locator
- Abstract, Methods and Findings, sentence 6
- How it applies
- Passage in the source that backs the claim.
- Retrieved
- Source effective date
- Not reported
- Source state
- Current
- Version fingerprint
0ace401db5d4f38c63e0108aab2fcbe976f2ad65e7e580d02f2def2833f917a0- Reuse basis
- PMC Open Access subset article under https://creativecommons.org/licenses/by/4.0/ (CC BY 4.0; the article's own licence element in the XML). The raw XML from NCBI E-utilities is committed unaltered in the atlas repository, which CC BY 4.0 allows with attribution. Short quotations under 25 words.
-
- Locator
- Abstract, Methods and Findings, sentence 7
- How it applies
- Passage in the source that backs the claim.
- Retrieved
- Source effective date
- Not reported
- Source state
- Current
- Version fingerprint
0ace401db5d4f38c63e0108aab2fcbe976f2ad65e7e580d02f2def2833f917a0- Reuse basis
- PMC Open Access subset article under https://creativecommons.org/licenses/by/4.0/ (CC BY 4.0; the article's own licence element in the XML). The raw XML from NCBI E-utilities is committed unaltered in the atlas repository, which CC BY 4.0 allows with attribution. Short quotations under 25 words.
-
- Locator
- Abstract, Methods and Findings, sentence 8
- How it applies
- Passage in the source that backs the claim.
- Retrieved
- Source effective date
- Not reported
- Source state
- Current
- Version fingerprint
0ace401db5d4f38c63e0108aab2fcbe976f2ad65e7e580d02f2def2833f917a0- Reuse basis
- PMC Open Access subset article under https://creativecommons.org/licenses/by/4.0/ (CC BY 4.0; the article's own licence element in the XML). The raw XML from NCBI E-utilities is committed unaltered in the atlas repository, which CC BY 4.0 allows with attribution. Short quotations under 25 words.
-
- Locator
- Abstract, Methods and Findings, sentence 10
- How it applies
- Passage in the source that backs the claim.
- Retrieved
- Source effective date
- Not reported
- Source state
- Current
- Version fingerprint
0ace401db5d4f38c63e0108aab2fcbe976f2ad65e7e580d02f2def2833f917a0- Reuse basis
- PMC Open Access subset article under https://creativecommons.org/licenses/by/4.0/ (CC BY 4.0; the article's own licence element in the XML). The raw XML from NCBI E-utilities is committed unaltered in the atlas repository, which CC BY 4.0 allows with attribution. Short quotations under 25 words.
Context Newborns screened for critical congenital heart defects with pulse oximetry, as a 2021 review describes the evidence
- Recipient
- Newborn
- Population
- Newborns screened by pulse oximetry for critical congenital heart defects (CCHDs). The review reports that, across the studies it included, data were provided for 457,202 newborns. It gives no gestational ages in the abstract.
- Care setting
- Not stated in the abstract
- Geography
- Not stated in the abstract
- Life stage
- The newborn period; the abstract gives no ages
- Gestational age
- Not reported
- Postnatal age
- Not reported
- Follow-up
- Not applicable (a review of recommendations and systematic reviews)
Eligibility: 1 inclusion criterion; exclusion criteria not stated
- Included
- Recommendations and systematic reviews on newborn pulse oximetry screening for CCHDs
- Excluded
- None stated
A 2021 review of recommendations and systematic reviews on newborn pulse oximetry screening for critical congenital heart defects (CCHDs) reports that current evidence supports consistent accuracy for detecting CCHDs when pulse oximetry screening is added to antenatal ultrasonography and clinical examination. It judges early diagnosis with pulse oximetry beneficial and cost-effective, the potential harms of false-positive tests not serious, and the consequences of missing CCHDs and other serious diseases detected by low blood oxygen serious. It reports that the site of testing (after the arterial duct only, or before and after it) had no significant effect on sensitivity or specificity.
Applies to Newborn Newborns screened for critical congenital heart defects with pulse oximetry, as a 2021 review describes the evidence
- Certainty method
- Not assessed
- Limitations
- A review of existing recommendations and systematic reviews with a literature search up to 13 August 2019, so it is a summary of other reviews. The abstract gives no sensitivity or specificity numbers, and beneficial and cost-effective is the review's judgement. It reports the number of newborns in the included studies (457,202) in its text and not as a finding of its own. Recipient: newborn.
- Last checked
- 2026-11-18
- Next evidence check due
- Not scheduled
Sources and citations (6)
-
Supports Newborn pulse oximetry screening for critical congenital heart defects
- Locator
- Abstract, sentence 2 (second part)
- How it applies
- Passage in the source that backs the claim.
- Retrieved
- Source effective date
- Not reported
- Source state
- Current
- Version fingerprint
55bc1e24764b48a0acf6e8207e84d1543617963b209e3782758c1f59b04347c2- Reuse basis
- PMC Open Access subset article under https://creativecommons.org/licenses/by/4.0/ (CC BY 4.0; the article's own licence element in the XML). The raw XML from NCBI E-utilities is committed unaltered in the atlas repository, which CC BY 4.0 allows with attribution. Short quotations under 25 words.
-
Supports Newborn pulse oximetry screening for critical congenital heart defects
- Locator
- Abstract, sentence 4 (second part)
- How it applies
- Passage in the source that backs the claim.
- Retrieved
- Source effective date
- Not reported
- Source state
- Current
- Version fingerprint
55bc1e24764b48a0acf6e8207e84d1543617963b209e3782758c1f59b04347c2- Reuse basis
- PMC Open Access subset article under https://creativecommons.org/licenses/by/4.0/ (CC BY 4.0; the article's own licence element in the XML). The raw XML from NCBI E-utilities is committed unaltered in the atlas repository, which CC BY 4.0 allows with attribution. Short quotations under 25 words.
-
Supports Newborn pulse oximetry screening for critical congenital heart defects
- Locator
- Abstract, sentence 5 (first part)
- How it applies
- Passage in the source that backs the claim.
- Retrieved
- Source effective date
- Not reported
- Source state
- Current
- Version fingerprint
55bc1e24764b48a0acf6e8207e84d1543617963b209e3782758c1f59b04347c2- Reuse basis
- PMC Open Access subset article under https://creativecommons.org/licenses/by/4.0/ (CC BY 4.0; the article's own licence element in the XML). The raw XML from NCBI E-utilities is committed unaltered in the atlas repository, which CC BY 4.0 allows with attribution. Short quotations under 25 words.
-
Supports Newborn pulse oximetry screening for critical congenital heart defects
- Locator
- Abstract, sentence 6
- How it applies
- Passage in the source that backs the claim.
- Retrieved
- Source effective date
- Not reported
- Source state
- Current
- Version fingerprint
55bc1e24764b48a0acf6e8207e84d1543617963b209e3782758c1f59b04347c2- Reuse basis
- PMC Open Access subset article under https://creativecommons.org/licenses/by/4.0/ (CC BY 4.0; the article's own licence element in the XML). The raw XML from NCBI E-utilities is committed unaltered in the atlas repository, which CC BY 4.0 allows with attribution. Short quotations under 25 words.
-
Supports Newborn pulse oximetry screening for critical congenital heart defects
- Locator
- Results, post-ductal versus pre- and post-ductal screening
- How it applies
- Passage in the source that backs the claim.
- Retrieved
- Source effective date
- Not reported
- Source state
- Current
- Version fingerprint
55bc1e24764b48a0acf6e8207e84d1543617963b209e3782758c1f59b04347c2- Reuse basis
- PMC Open Access subset article under https://creativecommons.org/licenses/by/4.0/ (CC BY 4.0; the article's own licence element in the XML). The raw XML from NCBI E-utilities is committed unaltered in the atlas repository, which CC BY 4.0 allows with attribution. Short quotations under 25 words.
-
Supports Newborn pulse oximetry screening for critical congenital heart defects
- Locator
- Body, included studies
- How it applies
- Passage in the source that backs the claim.
- Retrieved
- Source effective date
- Not reported
- Source state
- Current
- Version fingerprint
55bc1e24764b48a0acf6e8207e84d1543617963b209e3782758c1f59b04347c2- Reuse basis
- PMC Open Access subset article under https://creativecommons.org/licenses/by/4.0/ (CC BY 4.0; the article's own licence element in the XML). The raw XML from NCBI E-utilities is committed unaltered in the atlas repository, which CC BY 4.0 allows with attribution. Short quotations under 25 words.
Context People having their blood oxygen level checked, as a MedlinePlus medical test page describes them
- Recipient
- Mixed recipients
- Population
- A general public page about pulse oximetry for people of any age, with fingertip and home-use examples. It does not mention newborns.
- Care setting
- Not applicable (general public page)
- Geography
- Not stated in the source (a US public health information site)
- Life stage
- Not applicable; the page is not about a life stage
- Gestational age
- Not reported
- Postnatal age
- Not reported
- Follow-up
- Not applicable (general description)
Eligibility: 1 inclusion criterion; exclusion criteria not stated
- Included
- People having a blood oxygen level checked
- Excluded
- None stated
A MedlinePlus medical test page describes pulse oximetry as a test that uses a small, clip-like device called a pulse oximeter to measure oxygen levels in the blood, and as a painless and quick way to find out if the oxygen level is too low. It says there is no known risk to having it, and lists things that may cause inaccurate results, including dark nail polish, artificial nails, darker skin pigmentation, poor circulation, cold skin and tobacco use.
Applies to Mixed recipients People having their blood oxygen level checked, as a MedlinePlus medical test page describes them
- Certainty method
- Not assessed
- Limitations
- A general public page for people of any age, with fingertip and home-use examples; it does not mention newborns, and the factors that may cause inaccurate results are not tested in newborns on this page. Recipient: any age (general page, not about newborns).
- Last checked
- 2026-11-18
- Next evidence check due
- Not scheduled
Sources and citations (4)
-
Supports Pulse Oximetry (MedlinePlus medical test information)
- Locator
- What is pulse oximetry
- How it applies
- Passage in the source that backs the claim.
- Retrieved
- Source effective date
- 2024-09-12
- Source state
- Current
- Version fingerprint
0e7860547edffdfe2d7a9e8eb00ad97b4ce09543e9c780184f2c41b4844d0767- Reuse basis
- Medical test information on MedlinePlus is in the public domain, per the MedlinePlus page Linking to and Using Content from MedlinePlus (https://medlineplus.gov/about/using/usingcontent/, re-read 2026-11-08, page last updated October 26, 2022, sha256 0825c88f5bd1dade7c8693bad8035b0d1f25fd0b8c5053e488c618fb2b62e722). The page asks users to acknowledge MedlinePlus as the source. The page bytes are not committed; only the hash is kept. The page lists summaries on health topic pages and medical test information as public domain, and lists A.D.A.M. Medical Encyclopedia content, ASHP drug monographs and most images as copyrighted. The text quoted from this source is medical test information, not encyclopedia content, so it stays quote. Short quotations only.
-
Supports Pulse Oximetry (MedlinePlus medical test information)
- Locator
- Are there any risks to the test
- How it applies
- Passage in the source that backs the claim.
- Retrieved
- Source effective date
- 2024-09-12
- Source state
- Current
- Version fingerprint
0e7860547edffdfe2d7a9e8eb00ad97b4ce09543e9c780184f2c41b4844d0767- Reuse basis
- Medical test information on MedlinePlus is in the public domain, per the MedlinePlus page Linking to and Using Content from MedlinePlus (https://medlineplus.gov/about/using/usingcontent/, re-read 2026-11-08, page last updated October 26, 2022, sha256 0825c88f5bd1dade7c8693bad8035b0d1f25fd0b8c5053e488c618fb2b62e722). The page asks users to acknowledge MedlinePlus as the source. The page bytes are not committed; only the hash is kept. The page lists summaries on health topic pages and medical test information as public domain, and lists A.D.A.M. Medical Encyclopedia content, ASHP drug monographs and most images as copyrighted. The text quoted from this source is medical test information, not encyclopedia content, so it stays quote. Short quotations only.
-
Supports Pulse Oximetry (MedlinePlus medical test information)
- Locator
- Will I need to do anything to prepare
- How it applies
- Passage in the source that backs the claim.
- Retrieved
- Source effective date
- 2024-09-12
- Source state
- Current
- Version fingerprint
0e7860547edffdfe2d7a9e8eb00ad97b4ce09543e9c780184f2c41b4844d0767- Reuse basis
- Medical test information on MedlinePlus is in the public domain, per the MedlinePlus page Linking to and Using Content from MedlinePlus (https://medlineplus.gov/about/using/usingcontent/, re-read 2026-11-08, page last updated October 26, 2022, sha256 0825c88f5bd1dade7c8693bad8035b0d1f25fd0b8c5053e488c618fb2b62e722). The page asks users to acknowledge MedlinePlus as the source. The page bytes are not committed; only the hash is kept. The page lists summaries on health topic pages and medical test information as public domain, and lists A.D.A.M. Medical Encyclopedia content, ASHP drug monographs and most images as copyrighted. The text quoted from this source is medical test information, not encyclopedia content, so it stays quote. Short quotations only.
-
Supports Pulse Oximetry (MedlinePlus medical test information)
- Locator
- Will I need to do anything to prepare, list
- How it applies
- Passage in the source that backs the claim.
- Retrieved
- Source effective date
- 2024-09-12
- Source state
- Current
- Version fingerprint
0e7860547edffdfe2d7a9e8eb00ad97b4ce09543e9c780184f2c41b4844d0767- Reuse basis
- Medical test information on MedlinePlus is in the public domain, per the MedlinePlus page Linking to and Using Content from MedlinePlus (https://medlineplus.gov/about/using/usingcontent/, re-read 2026-11-08, page last updated October 26, 2022, sha256 0825c88f5bd1dade7c8693bad8035b0d1f25fd0b8c5053e488c618fb2b62e722). The page asks users to acknowledge MedlinePlus as the source. The page bytes are not committed; only the hash is kept. The page lists summaries on health topic pages and medical test information as public domain, and lists A.D.A.M. Medical Encyclopedia content, ASHP drug monographs and most images as copyrighted. The text quoted from this source is medical test information, not encyclopedia content, so it stays quote. Short quotations only.
Clinical links
Each link reports what one cited claim says about one population. "Evaluated for" means studied, not shown to work. These links do not show that anything is effective or right for you.
-
Claim, not a settled fact A 2021 review of pulse oximetry screening for critical congenital heart defects: Congenital heart disease
Applies to Newborn Newborns screened for critical congenital heart defects with pulse oximetry, as a 2021 review describes the evidence
A 2021 review of recommendations and systematic reviews on newborn pulse oximetry screening for critical congenital heart defects (CCHDs) reports that current evidence supports consistent accuracy for detecting CCHDs when pulse oximetry screening is added to antenatal ultrasonography and clinical examination. It judges early diagnosis with pulse oximetry beneficial and cost-effective, the potential harms of false-positive tests not serious, and the consequences of missing CCHDs and other serious diseases detected by low blood oxygen serious. It reports that the site of testing (after the arterial duct only, or before and after it) had no significant effect on sensitivity or specificity.Limitations. A review of existing recommendations and systematic reviews with a literature search up to 13 August 2019, so it is a summary of other reviews. The abstract gives no sensitivity or specificity numbers, and beneficial and cost-effective is the review's judgement. It reports the number of newborns in the included studies (457,202) in its text and not as a finding of its own. Recipient: newborn.
Record details
- Record id
maternal:neonatal-pulse-oximetry- Content hash
ddb1aac168077f4358d8f2abe2e4ec9adea1634960c8970e12abe6f435ce8378