Mechanical ventilators for newborns
A neonatal ventilator is a machine that pushes air into a newborn baby's lungs and lets it out again, through a tube in the windpipe. A 2020 review says that current ventilators are driven by microprocessors and can match the baby's own breaths, and that no single ventilator has been shown to be better than others. A 2025 review says that some all-purpose ventilators may not measure the very small breaths of the smallest babies accurately. This page names no brand and gives no settings.
Also called Neonatal ventilators · Neonatal ventilator
- Technology
- In this atlas
Overview
Technical summary
Technology page for the device class (MeSH D012122 Ventilators, Mechanical, broad: the descriptor covers ventilators for people of any age). The sourcing plan listed it as a product; it is written as a technology because no named product, model or regulatory decision was read, and a product record needs a model or formulation and an intended-use document. Sources read: the MeSH concept scope note and three CC BY 4.0 reviews (PMC12310796, PMC7025474, PMC9670452), with abstracts and selected paragraphs read. Not read: any manufacturer document, regulatory clearance or standard. The page states no setting, dose or manufacturer.
How it works
MeSH describes mechanical ventilators as mechanical devices used to produce or assist pulmonary ventilation. The 2025 review says that the basic modes depend on how the ventilator triggers, limits and cycles each breath, and names three high-frequency modalities: oscillatory, jet and flow interruption.
Classification codes
-
D012122Ventilators, Mechanical
Strengths
- A 2020 review says that current ventilators are driven by microprocessors and can synchronize the ventilator breath with the baby's own breath.
- A 2020 review says that newer ventilators have flow sensors that are very sensitive up to tidal volumes of 5 ml, and that volume-targeted ventilation relies heavily on the accurate measurement of expired tidal volume.
Limitations
- A 2020 review says that no single ventilator has been shown to be superior to others.
- A 2025 review says that older and universal ventilators may lack the accuracy needed for the tiny breaths of extremely low birth weight infants, which holds back volume-targeted ventilation.
Required infrastructure
- A neonatal intensive care unit; the 2025 review says that titration of ventilation requires an experienced clinician who judges at the bedside and uses real-time pulmonary graphics and monitoring.
Claims and evidence
Context A vocabulary definition (MeSH), not a study population
- Recipient
- No clinical recipient
- Population
- Not applicable: a controlled-vocabulary definition from the National Library of Medicine, with no population.
- Care setting
- Not applicable (vocabulary)
- Geography
- Not applicable (vocabulary)
- Life stage
- Not applicable (vocabulary)
- Follow-up
- Not applicable (vocabulary)
Eligibility: 1 inclusion criterion; exclusion criteria not stated
- Included
- Not applicable (vocabulary)
- Excluded
- None stated
The National Library of Medicine vocabulary MeSH defines mechanical ventilators as mechanical devices used to produce or assist pulmonary ventilation.
Applies to No clinical recipient A vocabulary definition (MeSH), not a study population
- Certainty method
- Not assessed
- Limitations
- A vocabulary definition for ventilators of any kind, for people of any age.
- Last checked
- 2026-11-18
- Next evidence check due
- Not scheduled
Sources and citations (1)
-
Supports MeSH concept M0018882 (preferred concept of descriptor D012122, Ventilators, Mechanical)
- Locator
- scopeNote.@value
- How it applies
- MeSH scope note (NLM, attribution given). A definition, not a clinical finding.
- Retrieved
- Source effective date
- Not reported
- Source state
- Current
- Version fingerprint
739be74017f3540139d6e71ecb316adda13052b5294c5142d888b57110731fea- Reuse basis
- The NLM web policies page (https://www.nlm.nih.gov/web_policies.html), re-read 2026-11-18 (sha256 c8b7026b5c3fcfd1b0cb88755077f6a99a5f2159eecd42b0e7d1a5ad9b30ec55, hash only), says works produced by the U.S. government are not subject to copyright protection in the United States and asks users to acknowledge NLM as the source. Source: National Library of Medicine. The raw bytes are committed in the atlas repository.
Context Extremely low birth weight (ELBW) infants who need ventilation, as a 2020 review describes them
- Recipient
- Newborn
- Population
- Extremely low birth weight (ELBW) newborns with respiratory failure who are on invasive ventilation, as the authors of a 2020 review describe them. The review notes that studies on ELBW babies specifically are few.
- Care setting
- Neonatal intensive care unit
- Geography
- Not stated in the abstract
- Life stage
- Extremely low birth weight newborns
- Gestational age
- Not reported
- Postnatal age
- Not reported
- Follow-up
- Not applicable (literature review)
Eligibility: 1 inclusion criterion; exclusion criteria not stated
- Included
- Extremely low birth weight newborns on invasive ventilation
- Excluded
- None stated
A 2020 review says that current ventilators are driven by microprocessors and have changed breathing support in extremely low birth weight babies by matching the machine breath to the baby's own breath. It says that high-frequency ventilators deliver tidal volumes smaller than the dead space, and that no single ventilator has been shown to be better than others.
Applies to Newborn Extremely low birth weight (ELBW) infants who need ventilation, as a 2020 review describes them
- Certainty method
- Not assessed
- Limitations
- Statements in a 2020 review. The review does not name or compare products and the atlas did not read a comparison trial.
- Last checked
- 2026-11-18
- Next evidence check due
- Not scheduled
Sources and citations (1)
-
- Locator
- Abstract, paragraph 1; Conclusion, paragraph 1; Volume Targeted (VTV) vs. Pressure Limited Ventilation (PLV) in ELBW infants, paragraph 8
- How it applies
- Full-text open-access review (CC BY 4.0), read in the PMC XML.
- Retrieved
- Source effective date
- Not reported
- Source state
- Current
- Version fingerprint
3a84f0e55f2379e33e61a6f73cb4091ed4999e0c2086b08eefa06e242627587b- Reuse basis
- PMC Open Access subset article under http://creativecommons.org/licenses/by/4.0/ (CC BY 4.0; the article's own permissions element). The raw XML from NCBI E-utilities is committed unaltered in the atlas repository, which the licence allows with attribution. Short quotations under 25 words.
Context Newborns in a neonatal intensive care unit who need invasive mechanical ventilation, as a 2025 review describes them
- Recipient
- Newborn
- Population
- Newborns whose breathing is supported through an endotracheal or tracheostomy tube in a neonatal intensive care unit, as the authors of a 2025 evidence-based review describe them. The review covers preterm and term babies and several lung and heart conditions.
- Care setting
- Neonatal intensive care unit
- Geography
- Not stated in the abstract (the review searched international databases)
- Life stage
- Newborn period, preterm and term
- Gestational age
- Not reported
- Postnatal age
- Not reported
- Follow-up
- Not applicable (literature review)
Eligibility: 1 inclusion criterion; exclusion criteria not stated
- Included
- Newborns who receive invasive mechanical ventilation
- Excluded
- None stated
The 2025 review says that volume-targeted ventilation is not used everywhere partly because older and universal ventilators may lack the accuracy needed to deliver the tiny breaths of extremely low birth weight infants, and that titration needs bedside assessment together with real-time pulmonary graphics and transcutaneous carbon dioxide or blood gas monitoring.
Applies to Newborn Newborns in a neonatal intensive care unit who need invasive mechanical ventilation, as a 2025 review describes them
- Certainty method
- Not assessed
- Limitations
- The authors give this as a probable reason in the discussion of volume-targeted ventilation, not as a measured finding. The 2020 review adds that newer ventilators have sensitive flow sensors, but that is the authors view too.
- Last checked
- 2026-11-18
- Next evidence check due
- Not scheduled
Sources and citations (1)
-
Supports Invasive Ventilation Strategies in Neonates
- Locator
- Volume-Targeted Ventilation, paragraph 1; Titration of Conventional Mechanical Ventilation, paragraph 1
- How it applies
- Full-text open-access review (CC BY 4.0), read in the PMC XML.
- Retrieved
- Source effective date
- Not reported
- Source state
- Current
- Version fingerprint
ebcc0173a783fa6e76356c0daa9c24b778eac8c795a50c2e3eab1192edc95ee8- Reuse basis
- PMC Open Access subset article under https://creativecommons.org/licenses/by/4.0/ (CC BY 4.0; the article's own permissions element). The raw XML from NCBI E-utilities is committed unaltered in the atlas repository, which the licence allows with attribution. Short quotations under 25 words.
The 2025 review says that the basic modes of mechanical ventilation depend on how the ventilator triggers a breath, limits it and cycles it into expiration, and that there are four basic modes of conventional ventilation: intermittent mandatory, synchronized intermittent mandatory, assist control and pressure support.
Applies to Newborn Newborns in a neonatal intensive care unit who need invasive mechanical ventilation, as a 2025 review describes them
- Certainty method
- Not assessed
- Limitations
- A description of modes for readers, not a comparison of machines. The review says the three synchronized modes differ in trigger, control and cycling; this page does not describe settings.
- Last checked
- 2026-11-18
- Next evidence check due
- Not scheduled
Sources and citations (1)
-
Supports Invasive Ventilation Strategies in Neonates
- Locator
- Conventional Mechanical Ventilation (CMV), paragraph 1; Conventional Mechanical Ventilation (CMV), paragraph 2
- How it applies
- Full-text open-access review (CC BY 4.0), read in the PMC XML.
- Retrieved
- Source effective date
- Not reported
- Source state
- Current
- Version fingerprint
ebcc0173a783fa6e76356c0daa9c24b778eac8c795a50c2e3eab1192edc95ee8- Reuse basis
- PMC Open Access subset article under https://creativecommons.org/licenses/by/4.0/ (CC BY 4.0; the article's own permissions element). The raw XML from NCBI E-utilities is committed unaltered in the atlas repository, which the licence allows with attribution. Short quotations under 25 words.
The 2025 review says that high-frequency ventilation gives very small breaths at rapid rates (up to 900 per minute), that the three main modalities are oscillatory, jet and flow interruption, and that the first two are commonly used in neonates. It describes the oscillator as a specialised ventilator that makes rapid small-amplitude oscillations.
Applies to Newborn Newborns in a neonatal intensive care unit who need invasive mechanical ventilation, as a 2025 review describes them
- Certainty method
- Not assessed
- Limitations
- Descriptive statements from a review. The review concludes that high-frequency ventilation should be kept for rescue support (see the ventilation page); this claim only says what the machines do.
- Last checked
- 2026-11-18
- Next evidence check due
- Not scheduled
Sources and citations (1)
-
Supports Invasive Ventilation Strategies in Neonates
- Locator
- High-Frequency Ventilation, paragraph 1; High-Frequency Oscillatory Ventilation, paragraph 1
- How it applies
- Full-text open-access review (CC BY 4.0), read in the PMC XML.
- Retrieved
- Source effective date
- Not reported
- Source state
- Current
- Version fingerprint
ebcc0173a783fa6e76356c0daa9c24b778eac8c795a50c2e3eab1192edc95ee8- Reuse basis
- PMC Open Access subset article under https://creativecommons.org/licenses/by/4.0/ (CC BY 4.0; the article's own permissions element). The raw XML from NCBI E-utilities is committed unaltered in the atlas repository, which the licence allows with attribution. Short quotations under 25 words.
Record details
- Record id
maternal:neonatal-ventilators- Content hash
f98a87054de683c141a965dc3570c267268deb92e5e8b923ce0d47fed631765d