Mechanical ventilators for newborns

A neonatal ventilator is a machine that pushes air into a newborn baby's lungs and lets it out again, through a tube in the windpipe. A 2020 review says that current ventilators are driven by microprocessors and can match the baby's own breaths, and that no single ventilator has been shown to be better than others. A 2025 review says that some all-purpose ventilators may not measure the very small breaths of the smallest babies accurately. This page names no brand and gives no settings.

Also called Neonatal ventilators · Neonatal ventilator

  • Technology
  • In this atlas

Overview

Technical summary

Technology page for the device class (MeSH D012122 Ventilators, Mechanical, broad: the descriptor covers ventilators for people of any age). The sourcing plan listed it as a product; it is written as a technology because no named product, model or regulatory decision was read, and a product record needs a model or formulation and an intended-use document. Sources read: the MeSH concept scope note and three CC BY 4.0 reviews (PMC12310796, PMC7025474, PMC9670452), with abstracts and selected paragraphs read. Not read: any manufacturer document, regulatory clearance or standard. The page states no setting, dose or manufacturer.

How it works

MeSH describes mechanical ventilators as mechanical devices used to produce or assist pulmonary ventilation. The 2025 review says that the basic modes depend on how the ventilator triggers, limits and cycles each breath, and names three high-frequency modalities: oscillatory, jet and flow interruption.

Classification codes

  • MeSH id.nlm.nih.gov linked data, retrieved 2026-11-18; descriptor lastUpdated 2023-05-31; no yearly MeSH release stated D012122 Ventilators, Mechanical

Strengths

  • A 2020 review says that current ventilators are driven by microprocessors and can synchronize the ventilator breath with the baby's own breath.
  • A 2020 review says that newer ventilators have flow sensors that are very sensitive up to tidal volumes of 5 ml, and that volume-targeted ventilation relies heavily on the accurate measurement of expired tidal volume.

Limitations

  • A 2020 review says that no single ventilator has been shown to be superior to others.
  • A 2025 review says that older and universal ventilators may lack the accuracy needed for the tiny breaths of extremely low birth weight infants, which holds back volume-targeted ventilation.

Required infrastructure

  • A neonatal intensive care unit; the 2025 review says that titration of ventilation requires an experienced clinician who judges at the bedside and uses real-time pulmonary graphics and monitoring.

Claims and evidence

Context A vocabulary definition (MeSH), not a study population

Recipient
No clinical recipient
Population
Not applicable: a controlled-vocabulary definition from the National Library of Medicine, with no population.
Care setting
Not applicable (vocabulary)
Geography
Not applicable (vocabulary)
Life stage
Not applicable (vocabulary)
Follow-up
Not applicable (vocabulary)
Eligibility: 1 inclusion criterion; exclusion criteria not stated
Included
  • Not applicable (vocabulary)
Excluded
None stated
  • Description Evidence stage not applicable Certainty not assessed

    The National Library of Medicine vocabulary MeSH defines mechanical ventilators as mechanical devices used to produce or assist pulmonary ventilation.

    Applies to No clinical recipient A vocabulary definition (MeSH), not a study population

    Certainty method
    Not assessed
    Limitations
    A vocabulary definition for ventilators of any kind, for people of any age.
    Last checked
    2026-11-18
    Next evidence check due
    Not scheduled
    Sources and citations (1)
    1. Supports MeSH concept M0018882 (preferred concept of descriptor D012122, Ventilators, Mechanical)

      Dataset · U.S. National Library of Medicine · Edition MeSH linked-data record of the preferred concept M0018882 of descriptor D012122, fetched 2026-11-18. The scope note is on this concept record, not on the descriptor record. The record states no date. · MeSH: M0018882

      Locator
      scopeNote.@value
      How it applies
      MeSH scope note (NLM, attribution given). A definition, not a clinical finding.
      Retrieved
      Source effective date
      Not reported
      Source state
      Current
      Version fingerprint
      739be74017f3540139d6e71ecb316adda13052b5294c5142d888b57110731fea
      Reuse basis
      The NLM web policies page (https://www.nlm.nih.gov/web_policies.html), re-read 2026-11-18 (sha256 c8b7026b5c3fcfd1b0cb88755077f6a99a5f2159eecd42b0e7d1a5ad9b30ec55, hash only), says works produced by the U.S. government are not subject to copyright protection in the United States and asks users to acknowledge NLM as the source. Source: National Library of Medicine. The raw bytes are committed in the atlas repository.

Context Extremely low birth weight (ELBW) infants who need ventilation, as a 2020 review describes them

Recipient
Newborn
Population
Extremely low birth weight (ELBW) newborns with respiratory failure who are on invasive ventilation, as the authors of a 2020 review describe them. The review notes that studies on ELBW babies specifically are few.
Geography
Not stated in the abstract
Life stage
Extremely low birth weight newborns
Gestational age
Not reported
Postnatal age
Not reported
Follow-up
Not applicable (literature review)
Eligibility: 1 inclusion criterion; exclusion criteria not stated
Included
  • Extremely low birth weight newborns on invasive ventilation
Excluded
None stated
  • Description Evidence synthesis Certainty not assessed

    A 2020 review says that current ventilators are driven by microprocessors and have changed breathing support in extremely low birth weight babies by matching the machine breath to the baby's own breath. It says that high-frequency ventilators deliver tidal volumes smaller than the dead space, and that no single ventilator has been shown to be better than others.

    Applies to Newborn Extremely low birth weight (ELBW) infants who need ventilation, as a 2020 review describes them

    Certainty method
    Not assessed
    Limitations
    Statements in a 2020 review. The review does not name or compare products and the atlas did not read a comparison trial.
    Last checked
    2026-11-18
    Next evidence check due
    Not scheduled
    Sources and citations (1)
    1. Supports Volume Targeted Ventilation and High Frequency Ventilation as the Primary Modes of Respiratory Support for ELBW Babies: What Does the Evidence Say?

      Publication · Frontiers in Pediatrics · Edition PMC Open Access subset full-text XML from NCBI E-utilities, retrieved 2026-11-18 (2020). Used: abstract and selected paragraphs. · PMCID: PMC7025474

      Locator
      Abstract, paragraph 1; Conclusion, paragraph 1; Volume Targeted (VTV) vs. Pressure Limited Ventilation (PLV) in ELBW infants, paragraph 8
      How it applies
      Full-text open-access review (CC BY 4.0), read in the PMC XML.
      Retrieved
      Source effective date
      Not reported
      Source state
      Current
      Version fingerprint
      3a84f0e55f2379e33e61a6f73cb4091ed4999e0c2086b08eefa06e242627587b
      Reuse basis
      PMC Open Access subset article under http://creativecommons.org/licenses/by/4.0/ (CC BY 4.0; the article's own permissions element). The raw XML from NCBI E-utilities is committed unaltered in the atlas repository, which the licence allows with attribution. Short quotations under 25 words.

Context Newborns in a neonatal intensive care unit who need invasive mechanical ventilation, as a 2025 review describes them

Recipient
Newborn
Population
Newborns whose breathing is supported through an endotracheal or tracheostomy tube in a neonatal intensive care unit, as the authors of a 2025 evidence-based review describe them. The review covers preterm and term babies and several lung and heart conditions.
Care setting
Neonatal intensive care unit
Geography
Not stated in the abstract (the review searched international databases)
Life stage
Newborn period, preterm and term
Gestational age
Not reported
Postnatal age
Not reported
Follow-up
Not applicable (literature review)
Eligibility: 1 inclusion criterion; exclusion criteria not stated
Included
Excluded
None stated
  • Description Evidence synthesis Certainty not assessed

    The 2025 review says that volume-targeted ventilation is not used everywhere partly because older and universal ventilators may lack the accuracy needed to deliver the tiny breaths of extremely low birth weight infants, and that titration needs bedside assessment together with real-time pulmonary graphics and transcutaneous carbon dioxide or blood gas monitoring.

    Applies to Newborn Newborns in a neonatal intensive care unit who need invasive mechanical ventilation, as a 2025 review describes them

    Certainty method
    Not assessed
    Limitations
    The authors give this as a probable reason in the discussion of volume-targeted ventilation, not as a measured finding. The 2020 review adds that newer ventilators have sensitive flow sensors, but that is the authors view too.
    Last checked
    2026-11-18
    Next evidence check due
    Not scheduled
    Sources and citations (1)
    1. Supports Invasive Ventilation Strategies in Neonates

      Publication · Indian Pediatrics · Edition PMC Open Access subset full-text XML from NCBI E-utilities, retrieved 2026-11-18 (2025). Used: abstract and selected paragraphs. · PMCID: PMC12310796

      Locator
      Volume-Targeted Ventilation, paragraph 1; Titration of Conventional Mechanical Ventilation, paragraph 1
      How it applies
      Full-text open-access review (CC BY 4.0), read in the PMC XML.
      Retrieved
      Source effective date
      Not reported
      Source state
      Current
      Version fingerprint
      ebcc0173a783fa6e76356c0daa9c24b778eac8c795a50c2e3eab1192edc95ee8
      Reuse basis
      PMC Open Access subset article under https://creativecommons.org/licenses/by/4.0/ (CC BY 4.0; the article's own permissions element). The raw XML from NCBI E-utilities is committed unaltered in the atlas repository, which the licence allows with attribution. Short quotations under 25 words.
  • Description Evidence synthesis Certainty not assessed

    The 2025 review says that the basic modes of mechanical ventilation depend on how the ventilator triggers a breath, limits it and cycles it into expiration, and that there are four basic modes of conventional ventilation: intermittent mandatory, synchronized intermittent mandatory, assist control and pressure support.

    Applies to Newborn Newborns in a neonatal intensive care unit who need invasive mechanical ventilation, as a 2025 review describes them

    Certainty method
    Not assessed
    Limitations
    A description of modes for readers, not a comparison of machines. The review says the three synchronized modes differ in trigger, control and cycling; this page does not describe settings.
    Last checked
    2026-11-18
    Next evidence check due
    Not scheduled
    Sources and citations (1)
    1. Supports Invasive Ventilation Strategies in Neonates

      Publication · Indian Pediatrics · Edition PMC Open Access subset full-text XML from NCBI E-utilities, retrieved 2026-11-18 (2025). Used: abstract and selected paragraphs. · PMCID: PMC12310796

      Locator
      Conventional Mechanical Ventilation (CMV), paragraph 1; Conventional Mechanical Ventilation (CMV), paragraph 2
      How it applies
      Full-text open-access review (CC BY 4.0), read in the PMC XML.
      Retrieved
      Source effective date
      Not reported
      Source state
      Current
      Version fingerprint
      ebcc0173a783fa6e76356c0daa9c24b778eac8c795a50c2e3eab1192edc95ee8
      Reuse basis
      PMC Open Access subset article under https://creativecommons.org/licenses/by/4.0/ (CC BY 4.0; the article's own permissions element). The raw XML from NCBI E-utilities is committed unaltered in the atlas repository, which the licence allows with attribution. Short quotations under 25 words.
  • Description Evidence synthesis Certainty not assessed

    The 2025 review says that high-frequency ventilation gives very small breaths at rapid rates (up to 900 per minute), that the three main modalities are oscillatory, jet and flow interruption, and that the first two are commonly used in neonates. It describes the oscillator as a specialised ventilator that makes rapid small-amplitude oscillations.

    Applies to Newborn Newborns in a neonatal intensive care unit who need invasive mechanical ventilation, as a 2025 review describes them

    Certainty method
    Not assessed
    Limitations
    Descriptive statements from a review. The review concludes that high-frequency ventilation should be kept for rescue support (see the ventilation page); this claim only says what the machines do.
    Last checked
    2026-11-18
    Next evidence check due
    Not scheduled
    Sources and citations (1)
    1. Supports Invasive Ventilation Strategies in Neonates

      Publication · Indian Pediatrics · Edition PMC Open Access subset full-text XML from NCBI E-utilities, retrieved 2026-11-18 (2025). Used: abstract and selected paragraphs. · PMCID: PMC12310796

      Locator
      High-Frequency Ventilation, paragraph 1; High-Frequency Oscillatory Ventilation, paragraph 1
      How it applies
      Full-text open-access review (CC BY 4.0), read in the PMC XML.
      Retrieved
      Source effective date
      Not reported
      Source state
      Current
      Version fingerprint
      ebcc0173a783fa6e76356c0daa9c24b778eac8c795a50c2e3eab1192edc95ee8
      Reuse basis
      PMC Open Access subset article under https://creativecommons.org/licenses/by/4.0/ (CC BY 4.0; the article's own permissions element). The raw XML from NCBI E-utilities is committed unaltered in the atlas repository, which the licence allows with attribution. Short quotations under 25 words.

Record details