# Mechanical ventilators for newborns

Technology · id `maternal:neonatal-ventilators` · Maternal & Newborn Atlas · release `8f5d5a5ed7ba4f08` (public)

> General information from published sources, not medical advice, and not reviewed by clinicians. Check the cited source before relying on anything here. Sponsors do not influence what this atlas says.

The atlas's own text and curation may be reused under Creative Commons Attribution-NonCommercial 4.0 International (CC BY-NC 4.0, https://creativecommons.org/licenses/by-nc/4.0/): credit the atlas, and do not use it commercially without a licence from the owner. See the atlas About page. Text quoted from a source stays under that source's own terms, given as the source's reuse\_basis. The atlas's reuse terms do not cover it. The downloads list each quotation with its source in third\_party\_quotations.

Also called: Neonatal ventilators; Neonatal ventilator

Browse, life stage: Newborn  
Browse, care area: Preterm and sick newborns  

## In plain language

A neonatal ventilator is a machine that pushes air into a newborn baby's lungs and lets it out again, through a tube in the windpipe. A 2020 review says that current ventilators are driven by microprocessors and can match the baby's own breaths, and that no single ventilator has been shown to be better than others. A 2025 review says that some all-purpose ventilators may not measure the very small breaths of the smallest babies accurately. This page names no brand and gives no settings.

## Technical summary

Technology page for the device class (MeSH D012122 Ventilators, Mechanical, broad: the descriptor covers ventilators for people of any age). The sourcing plan listed it as a product; it is written as a technology because no named product, model or regulatory decision was read, and a product record needs a model or formulation and an intended-use document. Sources read: the MeSH concept scope note and three CC BY 4.0 reviews (PMC12310796, PMC7025474, PMC9670452), with abstracts and selected paragraphs read. Not read: any manufacturer document, regulatory clearance or standard. The page states no setting, dose or manufacturer.

## Claims, with the population each applies to

## Context: A vocabulary definition (MeSH), not a study population

Recipient: No clinical recipient; Population: Not applicable: a controlled-vocabulary definition from the National Library of Medicine, with no population.; Care setting: Not applicable (vocabulary); Geography: Not applicable (vocabulary); Life stage: Not applicable (vocabulary); Follow-up: Not applicable (vocabulary)

### The National Library of Medicine vocabulary MeSH defines mechanical ventilators as mechanical devices used to produce or assist pulmonary ventilation.

- Category: Description
- Evidence stage: Evidence stage not applicable
- Certainty: Certainty not assessed
- Applies to: Recipient: No clinical recipient; Population: Not applicable: a controlled-vocabulary definition from the National Library of Medicine, with no population.; Care setting: Not applicable (vocabulary); Geography: Not applicable (vocabulary); Life stage: Not applicable (vocabulary); Follow-up: Not applicable (vocabulary)
- Limitations: A vocabulary definition for ventilators of any kind, for people of any age.
- Last checked: 2026-11-18
- Sources:
  1. Supports: MeSH concept M0018882 (preferred concept of descriptor D012122, Ventilators, Mechanical) (Dataset, edition MeSH linked-data record of the preferred concept M0018882 of descriptor D012122, fetched 2026-11-18. The scope note is on this concept record, not on the descriptor record. The record states no date.) <https://id.nlm.nih.gov/mesh/M0018882>
     Locator: scopeNote.@value. Retrieved 2026-11-18T21:43:43Z. Source state: Current. Version fingerprint: 739be74017f3540139d6e71ecb316adda13052b5294c5142d888b57110731fea.
     How it applies: MeSH scope note (NLM, attribution given). A definition, not a clinical finding.

## Context: Extremely low birth weight (ELBW) infants who need ventilation, as a 2020 review describes them

Recipient: Newborn; Population: Extremely low birth weight (ELBW) newborns with respiratory failure who are on invasive ventilation, as the authors of a 2020 review describe them. The review notes that studies on ELBW babies specifically are few.; Care setting: Neonatal intensive care unit; Geography: Not stated in the abstract; Life stage: Extremely low birth weight newborns; Gestational age: Not reported; Postnatal age: Not reported; Follow-up: Not applicable (literature review)

### A 2020 review says that current ventilators are driven by microprocessors and have changed breathing support in extremely low birth weight babies by matching the machine breath to the baby's own breath. It says that high-frequency ventilators deliver tidal volumes smaller than the dead space, and that no single ventilator has been shown to be better than others.

- Category: Description
- Evidence stage: Evidence synthesis
- Certainty: Certainty not assessed
- Applies to: Recipient: Newborn; Population: Extremely low birth weight (ELBW) newborns with respiratory failure who are on invasive ventilation, as the authors of a 2020 review describe them. The review notes that studies on ELBW babies specifically are few.; Care setting: Neonatal intensive care unit; Geography: Not stated in the abstract; Life stage: Extremely low birth weight newborns; Gestational age: Not reported; Postnatal age: Not reported; Follow-up: Not applicable (literature review)
- Limitations: Statements in a 2020 review. The review does not name or compare products and the atlas did not read a comparison trial.
- Last checked: 2026-11-18
- Sources:
  1. Supports: Volume Targeted Ventilation and High Frequency Ventilation as the Primary Modes of Respiratory Support for ELBW Babies: What Does the Evidence Say? (Publication, edition PMC Open Access subset full-text XML from NCBI E-utilities, retrieved 2026-11-18 (2020). Used: abstract and selected paragraphs.) <https://pmc.ncbi.nlm.nih.gov/articles/PMC7025474/>
     Locator: Abstract, paragraph 1; Conclusion, paragraph 1; Volume Targeted (VTV) vs. Pressure Limited Ventilation (PLV) in ELBW infants, paragraph 8. Retrieved 2026-11-18T21:42:19Z. Source state: Current. Version fingerprint: 3a84f0e55f2379e33e61a6f73cb4091ed4999e0c2086b08eefa06e242627587b.
     How it applies: Full-text open-access review (CC BY 4.0), read in the PMC XML.

## Context: Newborns in a neonatal intensive care unit who need invasive mechanical ventilation, as a 2025 review describes them

Recipient: Newborn; Population: Newborns whose breathing is supported through an endotracheal or tracheostomy tube in a neonatal intensive care unit, as the authors of a 2025 evidence-based review describe them. The review covers preterm and term babies and several lung and heart conditions.; Care setting: Neonatal intensive care unit; Geography: Not stated in the abstract (the review searched international databases); Life stage: Newborn period, preterm and term; Gestational age: Not reported; Postnatal age: Not reported; Follow-up: Not applicable (literature review)

### The 2025 review says that volume-targeted ventilation is not used everywhere partly because older and universal ventilators may lack the accuracy needed to deliver the tiny breaths of extremely low birth weight infants, and that titration needs bedside assessment together with real-time pulmonary graphics and transcutaneous carbon dioxide or blood gas monitoring.

- Category: Description
- Evidence stage: Evidence synthesis
- Certainty: Certainty not assessed
- Applies to: Recipient: Newborn; Population: Newborns whose breathing is supported through an endotracheal or tracheostomy tube in a neonatal intensive care unit, as the authors of a 2025 evidence-based review describe them. The review covers preterm and term babies and several lung and heart conditions.; Care setting: Neonatal intensive care unit; Geography: Not stated in the abstract (the review searched international databases); Life stage: Newborn period, preterm and term; Gestational age: Not reported; Postnatal age: Not reported; Follow-up: Not applicable (literature review)
- Limitations: The authors give this as a probable reason in the discussion of volume-targeted ventilation, not as a measured finding. The 2020 review adds that newer ventilators have sensitive flow sensors, but that is the authors view too.
- Last checked: 2026-11-18
- Sources:
  1. Supports: Invasive Ventilation Strategies in Neonates (Publication, edition PMC Open Access subset full-text XML from NCBI E-utilities, retrieved 2026-11-18 (2025). Used: abstract and selected paragraphs.) <https://pmc.ncbi.nlm.nih.gov/articles/PMC12310796/>
     Locator: Volume-Targeted Ventilation, paragraph 1; Titration of Conventional Mechanical Ventilation, paragraph 1. Retrieved 2026-11-18T21:42:17Z. Source state: Current. Version fingerprint: ebcc0173a783fa6e76356c0daa9c24b778eac8c795a50c2e3eab1192edc95ee8.
     How it applies: Full-text open-access review (CC BY 4.0), read in the PMC XML.

### The 2025 review says that the basic modes of mechanical ventilation depend on how the ventilator triggers a breath, limits it and cycles it into expiration, and that there are four basic modes of conventional ventilation: intermittent mandatory, synchronized intermittent mandatory, assist control and pressure support.

- Category: Description
- Evidence stage: Evidence synthesis
- Certainty: Certainty not assessed
- Applies to: Recipient: Newborn; Population: Newborns whose breathing is supported through an endotracheal or tracheostomy tube in a neonatal intensive care unit, as the authors of a 2025 evidence-based review describe them. The review covers preterm and term babies and several lung and heart conditions.; Care setting: Neonatal intensive care unit; Geography: Not stated in the abstract (the review searched international databases); Life stage: Newborn period, preterm and term; Gestational age: Not reported; Postnatal age: Not reported; Follow-up: Not applicable (literature review)
- Limitations: A description of modes for readers, not a comparison of machines. The review says the three synchronized modes differ in trigger, control and cycling; this page does not describe settings.
- Last checked: 2026-11-18
- Sources:
  1. Supports: Invasive Ventilation Strategies in Neonates (Publication, edition PMC Open Access subset full-text XML from NCBI E-utilities, retrieved 2026-11-18 (2025). Used: abstract and selected paragraphs.) <https://pmc.ncbi.nlm.nih.gov/articles/PMC12310796/>
     Locator: Conventional Mechanical Ventilation (CMV), paragraph 1; Conventional Mechanical Ventilation (CMV), paragraph 2. Retrieved 2026-11-18T21:42:17Z. Source state: Current. Version fingerprint: ebcc0173a783fa6e76356c0daa9c24b778eac8c795a50c2e3eab1192edc95ee8.
     How it applies: Full-text open-access review (CC BY 4.0), read in the PMC XML.

### The 2025 review says that high-frequency ventilation gives very small breaths at rapid rates (up to 900 per minute), that the three main modalities are oscillatory, jet and flow interruption, and that the first two are commonly used in neonates. It describes the oscillator as a specialised ventilator that makes rapid small-amplitude oscillations.

- Category: Description
- Evidence stage: Evidence synthesis
- Certainty: Certainty not assessed
- Applies to: Recipient: Newborn; Population: Newborns whose breathing is supported through an endotracheal or tracheostomy tube in a neonatal intensive care unit, as the authors of a 2025 evidence-based review describe them. The review covers preterm and term babies and several lung and heart conditions.; Care setting: Neonatal intensive care unit; Geography: Not stated in the abstract (the review searched international databases); Life stage: Newborn period, preterm and term; Gestational age: Not reported; Postnatal age: Not reported; Follow-up: Not applicable (literature review)
- Limitations: Descriptive statements from a review. The review concludes that high-frequency ventilation should be kept for rescue support (see the ventilation page); this claim only says what the machines do.
- Last checked: 2026-11-18
- Sources:
  1. Supports: Invasive Ventilation Strategies in Neonates (Publication, edition PMC Open Access subset full-text XML from NCBI E-utilities, retrieved 2026-11-18 (2025). Used: abstract and selected paragraphs.) <https://pmc.ncbi.nlm.nih.gov/articles/PMC12310796/>
     Locator: High-Frequency Ventilation, paragraph 1; High-Frequency Oscillatory Ventilation, paragraph 1. Retrieved 2026-11-18T21:42:17Z. Source state: Current. Version fingerprint: ebcc0173a783fa6e76356c0daa9c24b778eac8c795a50c2e3eab1192edc95ee8.
     How it applies: Full-text open-access review (CC BY 4.0), read in the PMC XML.

## Connected pages

- A common reason babies are ventilated (the 2025 review names respiratory distress): Respiratory distress syndrome in preterm babies (`maternal:neonatal-respiratory-distress-syndrome`)
- A device for breathing support without a tube: Continuous positive airway pressure (CPAP) for newborns (`maternal:nasal-cpap`)
- Monitoring used beside the ventilator in the 2025 review: Transcutaneous blood gas monitoring in neonates (`maternal:transcutaneous-blood-gas-monitoring-neonates`)
- Filed under this front: Newborn support: warmth, breathing, feeding (`maternal:front-newborn-support`)
- Used by: Invasive mechanical ventilation for newborns (`maternal:neonatal-mechanical-ventilation`)

## Provenance

Content hash: f98a87054de683c141a965dc3570c267268deb92e5e8b923ce0d47fed631765d
Last substantive edit: 2026-11-18
Evidence last checked: 2026-11-18