Newborn blood glucose monitoring
Newborn blood glucose monitoring means measuring a baby's blood sugar to look for low levels (neonatal hypoglycaemia). This page rests on a 2023 review of six guidelines, a 2020 mini review, a 2006 neonatal unit study of a bedside glucose meter and a small study of a continuous glucose sensor in very small preterm babies. The 2023 review reports that all but one of the issuing societies recommend screening babies who are at high risk or have symptoms, but that the screening approaches differ and the glucose level that defines the condition is not uniform. In the 2006 study, one bedside meter model gave values that differed from a laboratory analyser, and its authors call it useful only as a screening device. This page gives no glucose thresholds and no testing schedule.
Also called Neonatal glucose screening · Neonatal glucose measurement · Glucose testing in newborns
- Intervention
- In this atlas
Overview
Technical summary
Test (intervention, subtype test). Sources read: abstracts of a 2023 descriptive review of guidelines (PMC10378472, Children, CC BY 4.0), a 2020 mini review (PMC7820332, Frontiers in Pediatrics, CC BY 4.0), a 2006 prospective NICU (neonatal intensive care unit) study of the Abbott PCx meter (PMC1660538, BMC Pediatrics, CC BY 2.0) and a small validation study of a continuous glucose monitoring system in very low birth weight neonates on parenteral nutrition (PMC6106728, Italian Journal of Pediatrics, CC BY 4.0). MeSH D001786 is the blood glucose substance and is recorded as a broad identifier only. The MedlinePlus blood glucose topic is about diabetes in adults and is not used. No glucose cut-off, schedule or treatment is stated, and none of the guideline texts was read (only the review's synthesis). Recipient: newborn.
About this intervention
- Type
- Test
- Mechanism
- Not reported
Classification codes
-
D001786Blood Glucose
Claims and evidence
Context Newborn babies at risk of low blood glucose, or with signs of it, as six guidelines describe them (a 2023 review of guidelines)
- Recipient
- Newborn
- Population
- Newborn infants who have no symptoms but are at high risk of low blood glucose, and newborn infants who have symptoms, as each reviewed guideline defines them. The abstract gives no further definition.
- Care setting
- Not stated in the abstract
- Geography
- Not stated as a single place; the abstract names guideline issuers from North America, Europe and Australia
- Life stage
- Newborn period; the abstract gives no age bounds
- Gestational age
- Not reported
- Postnatal age
- Not reported
- Follow-up
- Not applicable (a review of guidelines)
Eligibility: 1 inclusion criterion; exclusion criteria not stated
- Included
- Recent influential guidelines on diagnosis, screening, prevention and management of neonatal hypoglycaemia from six issuing bodies
- Excluded
- None stated
A 2023 descriptive review compared six influential guidelines on neonatal hypoglycaemia (low blood glucose in the newborn). It reports that all of the issuing societies except the Pediatric Endocrine Society recommend screening for it in newborns who are at high risk but have no symptoms, and in newborns who have symptoms, but that the guidelines do not give consistent screening approaches. It also reports that the guidelines say the diagnosis should be confirmed by a laboratory measurement of blood glucose, although treatment should not be delayed until the result is available, and that the definition is not uniform: it is generally agreed that a single glucose value cannot define the condition.
Applies to Newborn Newborn babies at risk of low blood glucose, or with signs of it, as six guidelines describe them (a 2023 review of guidelines)
- Certainty method
- Not assessed
- Limitations
- A review of guidelines, not of trials: it reports what the guidelines say, not whether screening improves outcomes. The abstract gives no glucose values, ages or schedules, so this page states none. Only the review abstract was read; no guideline text was read, and later guideline editions were not checked. Recipient: newborn.
- Last checked
- 2026-11-18
- Next evidence check due
- Not scheduled
Sources and citations (4)
-
Supports Diagnosis and Management of Neonatal Hypoglycemia: A Comprehensive Review of Guidelines
- Locator
- Abstract, sentence 6
- How it applies
- Passage in the source that backs the claim.
- Retrieved
- Source effective date
- Not reported
- Source state
- Current
- Version fingerprint
5ac180a1db0c0e1b998f9e14ea726f3cff783ae6bb9799ce0350d6b8929455ad- Reuse basis
- PMC Open Access subset article under https://creativecommons.org/licenses/by/4.0/ (CC BY 4.0; the article's own licence element in the XML). The raw XML from NCBI E-utilities is committed unaltered in the atlas repository, which CC BY 4.0 allows with attribution. Short quotations under 25 words.
-
Supports Diagnosis and Management of Neonatal Hypoglycemia: A Comprehensive Review of Guidelines
- Locator
- Abstract, sentence 7
- How it applies
- Passage in the source that backs the claim.
- Retrieved
- Source effective date
- Not reported
- Source state
- Current
- Version fingerprint
5ac180a1db0c0e1b998f9e14ea726f3cff783ae6bb9799ce0350d6b8929455ad- Reuse basis
- PMC Open Access subset article under https://creativecommons.org/licenses/by/4.0/ (CC BY 4.0; the article's own licence element in the XML). The raw XML from NCBI E-utilities is committed unaltered in the atlas repository, which CC BY 4.0 allows with attribution. Short quotations under 25 words.
-
Supports Diagnosis and Management of Neonatal Hypoglycemia: A Comprehensive Review of Guidelines
- Locator
- Abstract, sentence 8
- How it applies
- Passage in the source that backs the claim.
- Retrieved
- Source effective date
- Not reported
- Source state
- Current
- Version fingerprint
5ac180a1db0c0e1b998f9e14ea726f3cff783ae6bb9799ce0350d6b8929455ad- Reuse basis
- PMC Open Access subset article under https://creativecommons.org/licenses/by/4.0/ (CC BY 4.0; the article's own licence element in the XML). The raw XML from NCBI E-utilities is committed unaltered in the atlas repository, which CC BY 4.0 allows with attribution. Short quotations under 25 words.
-
Supports Diagnosis and Management of Neonatal Hypoglycemia: A Comprehensive Review of Guidelines
- Locator
- Abstract, sentence 3
- How it applies
- Passage in the source that backs the claim.
- Retrieved
- Source effective date
- Not reported
- Source state
- Current
- Version fingerprint
5ac180a1db0c0e1b998f9e14ea726f3cff783ae6bb9799ce0350d6b8929455ad- Reuse basis
- PMC Open Access subset article under https://creativecommons.org/licenses/by/4.0/ (CC BY 4.0; the article's own licence element in the XML). The raw XML from NCBI E-utilities is committed unaltered in the atlas repository, which CC BY 4.0 allows with attribution. Short quotations under 25 words.
Context Newborn babies at risk of transient low blood glucose, as a 2020 mini review describes them
- Recipient
- Newborn
- Population
- Babies at risk of transient neonatal hypoglycaemia. The abstract says the highest-risk groups are well defined and does not list them.
- Care setting
- Not stated in the abstract
- Geography
- Not stated in the abstract
- Life stage
- Newborn period; the abstract gives no age bounds
- Gestational age
- Not reported
- Postnatal age
- Not reported
- Follow-up
- Not applicable (a mini review)
Eligibility: 1 inclusion criterion; exclusion criteria not stated
- Included
- Recent advances and current thinking on clinical aspects of transient neonatal hypoglycaemia
- Excluded
- None stated
A 2020 mini review says the groups of babies at highest risk of low blood glucose are well defined, but the best frequency and duration of screening, and the glucose level at which treatment would prevent brain injury, remain uncertain. It says continuous interstitial glucose monitoring in a research setting gives useful information about the duration, frequency and severity of low glucose, and that it is unknown whether continuous monitoring brings clinical benefit or harm. It says oral dextrose gel is increasingly being recommended as a first-line treatment. Its conclusion is that current management of neonatal hypoglycaemia lacks a reliable evidence base.
Applies to Newborn Newborn babies at risk of transient low blood glucose, as a 2020 mini review describes them
- Certainty method
- Not assessed
- Limitations
- A mini review: it summarises recent advances and gives no pooled estimate. The abstract does not list the high-risk groups. Recipient: newborn.
- Last checked
- 2026-11-18
- Next evidence check due
- Not scheduled
Sources and citations (4)
-
Supports Clinical Aspects of Neonatal Hypoglycemia: A Mini Review
- Locator
- Abstract, Results
- How it applies
- Passage in the source that backs the claim.
- Retrieved
- Source effective date
- Not reported
- Source state
- Current
- Version fingerprint
fd8eb901860d296b4b12cb4fbc428fe02ae0a8900b97bd072589e6061328844b- Reuse basis
- PMC Open Access subset article under https://creativecommons.org/licenses/by/4.0/ (CC BY 4.0; the article's own licence element in the XML). The raw XML from NCBI E-utilities is committed unaltered in the atlas repository, which CC BY 4.0 allows with attribution. Short quotations under 25 words.
-
Supports Clinical Aspects of Neonatal Hypoglycemia: A Mini Review
- Locator
- Abstract, Results (continuous monitoring)
- How it applies
- Passage in the source that backs the claim.
- Retrieved
- Source effective date
- Not reported
- Source state
- Current
- Version fingerprint
fd8eb901860d296b4b12cb4fbc428fe02ae0a8900b97bd072589e6061328844b- Reuse basis
- PMC Open Access subset article under https://creativecommons.org/licenses/by/4.0/ (CC BY 4.0; the article's own licence element in the XML). The raw XML from NCBI E-utilities is committed unaltered in the atlas repository, which CC BY 4.0 allows with attribution. Short quotations under 25 words.
-
Supports Clinical Aspects of Neonatal Hypoglycemia: A Mini Review
- Locator
- Abstract, Results (treatment)
- How it applies
- Passage in the source that backs the claim.
- Retrieved
- Source effective date
- Not reported
- Source state
- Current
- Version fingerprint
fd8eb901860d296b4b12cb4fbc428fe02ae0a8900b97bd072589e6061328844b- Reuse basis
- PMC Open Access subset article under https://creativecommons.org/licenses/by/4.0/ (CC BY 4.0; the article's own licence element in the XML). The raw XML from NCBI E-utilities is committed unaltered in the atlas repository, which CC BY 4.0 allows with attribution. Short quotations under 25 words.
-
Supports Clinical Aspects of Neonatal Hypoglycemia: A Mini Review
- Locator
- Abstract, Conclusion
- How it applies
- Passage in the source that backs the claim.
- Retrieved
- Source effective date
- Not reported
- Source state
- Current
- Version fingerprint
fd8eb901860d296b4b12cb4fbc428fe02ae0a8900b97bd072589e6061328844b- Reuse basis
- PMC Open Access subset article under https://creativecommons.org/licenses/by/4.0/ (CC BY 4.0; the article's own licence element in the XML). The raw XML from NCBI E-utilities is committed unaltered in the atlas repository, which CC BY 4.0 allows with attribution. Short quotations under 25 words.
Context Patients in one neonatal intensive care unit who had a plasma glucose test ordered (475 samples from 132 patients)
- Recipient
- Newborn
- Population
- All patients in the neonatal intensive care unit who had a plasma glucose test ordered were eligible. 475 samples from 132 patients were analysed. The abstract does not give their gestational ages or birth weights.
- Care setting
- A neonatal intensive care unit
- Geography
- Not stated in the abstract
- Life stage
- Neonatal intensive care patients; the abstract gives no ages
- Gestational age
- Not reported
- Postnatal age
- Not reported
- Follow-up
- Single bedside measurement per sample, compared with a laboratory measurement of the same sample
Eligibility: 1 inclusion criterion; exclusion criteria not stated
- Included
- Neonatal intensive care patients with a plasma glucose test ordered
- Excluded
- None stated
In a prospective 2006 study of 475 samples from 132 patients in one neonatal intensive care unit, glucose values from a bedside point-of-care meter (the Abbott PCx) differed from laboratory analyser values: bedside values averaged 4.9% higher, and values from the same model of meter run in the laboratory averaged 5.2% lower. Haematocrit explained about 30% of the variance between meter and analyser, with a greater effect at glucose of 4.0 mmol/L or less. The authors conclude that the meter is useful only as a screening device for neonatal hypoglycaemia and that a screening cut-off value must be established.
Applies to Newborn Patients in one neonatal intensive care unit who had a plasma glucose test ordered (475 samples from 132 patients)
- Certainty method
- Not assessed
- Limitations
- One meter model in one unit, published in 2006: other meters and later models may differ. The abstract gives no gestational ages or birth weights. It also gives a cut-off value for confirmatory testing, which this page does not repeat. Recipient: newborn.
- Last checked
- 2026-11-18
- Next evidence check due
- Not scheduled
Sources and citations (4)
-
- Locator
- Abstract, Results, sentence 1
- How it applies
- Passage in the source that backs the claim.
- Retrieved
- Source effective date
- Not reported
- Source state
- Current
- Version fingerprint
2fd57e43107aa34a7d4c8a39bcbc54ccc1e7a5b7d59848a8c3497d21d430c7b9- Reuse basis
- PMC Open Access subset article under https://creativecommons.org/licenses/by/2.0/ (CC BY 2.0; the article's own licence element in the XML). The raw XML from NCBI E-utilities is committed unaltered in the atlas repository, which CC BY 2.0 allows with attribution. Short quotations under 25 words.
-
- Locator
- Abstract, Results, sentence 2
- How it applies
- Passage in the source that backs the claim.
- Retrieved
- Source effective date
- Not reported
- Source state
- Current
- Version fingerprint
2fd57e43107aa34a7d4c8a39bcbc54ccc1e7a5b7d59848a8c3497d21d430c7b9- Reuse basis
- PMC Open Access subset article under https://creativecommons.org/licenses/by/2.0/ (CC BY 2.0; the article's own licence element in the XML). The raw XML from NCBI E-utilities is committed unaltered in the atlas repository, which CC BY 2.0 allows with attribution. Short quotations under 25 words.
-
- Locator
- Abstract, Results, sentence 4
- How it applies
- Passage in the source that backs the claim.
- Retrieved
- Source effective date
- Not reported
- Source state
- Current
- Version fingerprint
2fd57e43107aa34a7d4c8a39bcbc54ccc1e7a5b7d59848a8c3497d21d430c7b9- Reuse basis
- PMC Open Access subset article under https://creativecommons.org/licenses/by/2.0/ (CC BY 2.0; the article's own licence element in the XML). The raw XML from NCBI E-utilities is committed unaltered in the atlas repository, which CC BY 2.0 allows with attribution. Short quotations under 25 words.
-
- Locator
- Abstract, Conclusion, sentence 2
- How it applies
- Passage in the source that backs the claim.
- Retrieved
- Source effective date
- Not reported
- Source state
- Current
- Version fingerprint
2fd57e43107aa34a7d4c8a39bcbc54ccc1e7a5b7d59848a8c3497d21d430c7b9- Reuse basis
- PMC Open Access subset article under https://creativecommons.org/licenses/by/2.0/ (CC BY 2.0; the article's own licence element in the XML). The raw XML from NCBI E-utilities is committed unaltered in the atlas repository, which CC BY 2.0 allows with attribution. Short quotations under 25 words.
Context Very low birth weight newborns who needed parenteral nutrition (23 babies, continuous glucose sensor for at least 72 hours)
- Recipient
- Newborn
- Population
- 23 very low birth weight neonates (9 male) enrolled within 24 hours of birth, median gestational age 28 weeks (range 23 to 30) and median birth weight 860 g (range 500 to 1092), who needed parenteral nutrition.
- Care setting
- Neonatal intensive care (the abstract describes neonatal intensive care use and does not name the unit)
- Geography
- Not stated in the abstract
- Life stage
- Enrolled within 24 hours of birth; sensor kept for at least 72 hours
- Gestational age
- Not reported
- Postnatal age
- Not reported
- Follow-up
- At least 72 hours of sensor monitoring with calibration every 12 hours
Eligibility: 1 inclusion criterion; exclusion criteria not stated
- Included
- Very low birth weight neonates enrolled within 24 hours from birth who needed parenteral nutrition
- Excluded
- None stated
In a study of 23 very low birth weight newborns who needed parenteral nutrition (median gestational age 28 weeks, range 23 to 30; median birth weight 860 g, range 500 to 1092), a subcutaneous continuous glucose monitoring sensor (an ENLITE sensor) was kept in place for at least 72 hours and compared with a glucometer on 299 paired values. The authors report that the Modified Clarke Error Grid criteria for clinical significance were met, that the system detected 74 episodes of hypoglycaemia and 33 of hyperglycaemia, and that it is safe and accurate.
Applies to Newborn Very low birth weight newborns who needed parenteral nutrition (23 babies, continuous glucose sensor for at least 72 hours)
- Certainty method
- Not assessed
- Limitations
- A small study of 23 babies. The abstract does not name the centre or country and reports no clinical outcome. Safe and accurate is the authors' conclusion. A 2020 mini review on this page says it remains unknown whether continuous monitoring brings clinical benefit or harm. Recipient: newborn.
- Last checked
- 2026-11-18
- Next evidence check due
- Not scheduled
Sources and citations (7)
-
- Locator
- Abstract, Background, sentence 1
- How it applies
- Passage in the source that backs the claim.
- Retrieved
- Source effective date
- Not reported
- Source state
- Current
- Version fingerprint
3e5510e3bc5b5464c61253df1a4cb46c7a354bf69e8c3a50c7b61fb4edf52796- Reuse basis
- PMC Open Access subset article under https://creativecommons.org/licenses/by/4.0/ (CC BY 4.0; the article's own licence element in the XML). The raw XML from NCBI E-utilities is committed unaltered in the atlas repository, which CC BY 4.0 allows with attribution. Short quotations under 25 words.
-
- Locator
- Abstract, Results, sentence 1
- How it applies
- Passage in the source that backs the claim.
- Retrieved
- Source effective date
- Not reported
- Source state
- Current
- Version fingerprint
3e5510e3bc5b5464c61253df1a4cb46c7a354bf69e8c3a50c7b61fb4edf52796- Reuse basis
- PMC Open Access subset article under https://creativecommons.org/licenses/by/4.0/ (CC BY 4.0; the article's own licence element in the XML). The raw XML from NCBI E-utilities is committed unaltered in the atlas repository, which CC BY 4.0 allows with attribution. Short quotations under 25 words.
-
- Locator
- Abstract, Results, sentence 2
- How it applies
- Passage in the source that backs the claim.
- Retrieved
- Source effective date
- Not reported
- Source state
- Current
- Version fingerprint
3e5510e3bc5b5464c61253df1a4cb46c7a354bf69e8c3a50c7b61fb4edf52796- Reuse basis
- PMC Open Access subset article under https://creativecommons.org/licenses/by/4.0/ (CC BY 4.0; the article's own licence element in the XML). The raw XML from NCBI E-utilities is committed unaltered in the atlas repository, which CC BY 4.0 allows with attribution. Short quotations under 25 words.
-
- Locator
- Abstract, Methods, sentence 2
- How it applies
- Passage in the source that backs the claim.
- Retrieved
- Source effective date
- Not reported
- Source state
- Current
- Version fingerprint
3e5510e3bc5b5464c61253df1a4cb46c7a354bf69e8c3a50c7b61fb4edf52796- Reuse basis
- PMC Open Access subset article under https://creativecommons.org/licenses/by/4.0/ (CC BY 4.0; the article's own licence element in the XML). The raw XML from NCBI E-utilities is committed unaltered in the atlas repository, which CC BY 4.0 allows with attribution. Short quotations under 25 words.
-
- Locator
- Abstract, Results, sentence 3
- How it applies
- Passage in the source that backs the claim.
- Retrieved
- Source effective date
- Not reported
- Source state
- Current
- Version fingerprint
3e5510e3bc5b5464c61253df1a4cb46c7a354bf69e8c3a50c7b61fb4edf52796- Reuse basis
- PMC Open Access subset article under https://creativecommons.org/licenses/by/4.0/ (CC BY 4.0; the article's own licence element in the XML). The raw XML from NCBI E-utilities is committed unaltered in the atlas repository, which CC BY 4.0 allows with attribution. Short quotations under 25 words.
-
- Locator
- Abstract, Results, sentence 4
- How it applies
- Passage in the source that backs the claim.
- Retrieved
- Source effective date
- Not reported
- Source state
- Current
- Version fingerprint
3e5510e3bc5b5464c61253df1a4cb46c7a354bf69e8c3a50c7b61fb4edf52796- Reuse basis
- PMC Open Access subset article under https://creativecommons.org/licenses/by/4.0/ (CC BY 4.0; the article's own licence element in the XML). The raw XML from NCBI E-utilities is committed unaltered in the atlas repository, which CC BY 4.0 allows with attribution. Short quotations under 25 words.
-
- Locator
- Abstract, Conclusions, sentence 1
- How it applies
- Passage in the source that backs the claim.
- Retrieved
- Source effective date
- Not reported
- Source state
- Current
- Version fingerprint
3e5510e3bc5b5464c61253df1a4cb46c7a354bf69e8c3a50c7b61fb4edf52796- Reuse basis
- PMC Open Access subset article under https://creativecommons.org/licenses/by/4.0/ (CC BY 4.0; the article's own licence element in the XML). The raw XML from NCBI E-utilities is committed unaltered in the atlas repository, which CC BY 4.0 allows with attribution. Short quotations under 25 words.
Clinical links
Each link reports what one cited claim says about one population. "Evaluated for" means studied, not shown to work. These links do not show that anything is effective or right for you.
-
Claim, not a settled fact A 2023 review of guidelines on screening newborns for low blood glucose: Low blood sugar in newborns (neonatal hypoglycaemia)
Applies to Newborn Newborn babies at risk of low blood glucose, or with signs of it, as six guidelines describe them (a 2023 review of guidelines)
A 2023 descriptive review compared six influential guidelines on neonatal hypoglycaemia (low blood glucose in the newborn). It reports that all of the issuing societies except the Pediatric Endocrine Society recommend screening for it in newborns who are at high risk but have no symptoms, and in newborns who have symptoms, but that the guidelines do not give consistent screening approaches. It also reports that the guidelines say the diagnosis should be confirmed by a laboratory measurement of blood glucose, although treatment should not be delayed until the result is available, and that the definition is not uniform: it is generally agreed that a single glucose value cannot define the condition.Limitations. A review of guidelines, not of trials: it reports what the guidelines say, not whether screening improves outcomes. The abstract gives no glucose values, ages or schedules, so this page states none. Only the review abstract was read; no guideline text was read, and later guideline editions were not checked. Recipient: newborn.