# Vacuum extractor devices

Product · id `maternal:vacuum-extractor-devices` · Maternal & Newborn Atlas · release `8f5d5a5ed7ba4f08` (public)

> General information from published sources, not medical advice, and not reviewed by clinicians. Check the cited source before relying on anything here. Sponsors do not influence what this atlas says.

The atlas's own text and curation may be reused under Creative Commons Attribution-NonCommercial 4.0 International (CC BY-NC 4.0, https://creativecommons.org/licenses/by-nc/4.0/): credit the atlas, and do not use it commercially without a licence from the owner. See the atlas About page. Text quoted from a source stays under that source's own terms, given as the source's reuse\_basis. The atlas's reuse terms do not cover it. The downloads list each quotation with its source in third\_party\_quotations.

Also called: Vacuum-assisted delivery devices; Fetal vacuum extractors; Vacuum cups for assisted birth

Browse, life stage: Childbirth  
Browse, care area: Labour and birth  

## In plain language

A vacuum extractor is a suction cup device used to help with delivery. US federal rules identify the fetal vacuum extractor as a device that applies traction to the fetal head through a suction cup attached to the scalp and place the generic type in Class II. Reviews report that vacuum extraction carries a lower risk of maternal perineal trauma than forceps but higher failure rates and more scalp injury in babies, that soft and rigid cups trade failure against scalp injury, and that the evidence behind device design is thin. This page describes the device type, not any brand, and gives no advice on use.

## Technical summary

Product, subtype device; a generic device type, not a single brand or model. FDA product code HDB (Extractor, Vacuum, Fetal), 21 CFR 884.4340, device class 2 (openFDA device classification record fetched 2026-11-19; eCFR text for 2026-11-01). Two open-access reviews were read: a 2020 engineering review (PMC7780266, CC BY 4.0) and a 2026 expert commentary that proposes an ultrasound method (PMC13024969, CC BY 4.0). Their statements about other studies cite works not read. No manufacturer, brand or 510(k) summary was read, so no other product record is linked. The operative vaginal birth record holds the care-pathway claims and is linked, not repeated. The commentary's proposed method and any technique, traction, timing or gestational-age rules are left out.

## Claims, with the population each applies to

## Context: Babies born with a vacuum extractor (reviews 2020 and 2026)

Recipient: Newborn; Population: Babies delivered with a vacuum cup on the scalp, as described by the reviews. Figures are background statements from other studies the atlas has not read.; Care setting: Hospital labour wards.; Geography: Not stated in the source; Life stage: Birth and the first days after birth; the sources give no age band; Gestational age: Not reported; Postnatal age: Not reported; Follow-up: Not applicable (literature review)

### Reviews report scalp and head injuries in babies after vacuum birth. The 2026 commentary lists cephalohematoma and intracranial hemorrhage among the complications of vacuum extraction and says that, compared with forceps, vacuum extraction carries a higher risk of neonatal scalp injury. The 2020 review gives subgaleal haematomas in approximately 6 in 10,000 vacuum deliveries, and the commentary cites a retrospective cohort study in which cup detachment went with higher rates of subgaleal hematoma (8.9% compared with 3.5%).

- Category: Safety
- Evidence stage: Evidence synthesis
- Certainty: Certainty not assessed
- Applies to: Recipient: Newborn; Population: Babies delivered with a vacuum cup on the scalp, as described by the reviews. Figures are background statements from other studies the atlas has not read.; Care setting: Hospital labour wards.; Geography: Not stated in the source; Life stage: Birth and the first days after birth; the sources give no age band; Gestational age: Not reported; Postnatal age: Not reported; Follow-up: Not applicable (literature review)
- Limitations: Background statements that cite studies the atlas did not read; the figures come from different populations and definitions and are not combined. Certainty is not assessed here.
- Last checked: 2026-11-19
- Sources:
  1. Supports: Intrapartum Ultrasound in Vacuum Operative Delivery: A Comprehensive Review and Proposal of the Novel Ultrasound Flexion Point Method (Publication, edition PMC Open Access subset full-text XML from NCBI E-utilities, retrieved 2026-11-19 (2026). An expert commentary and technical proposal. Used: the abstract, the introduction and the instrument-selection discussion. The proposed ultrasound method is not described in the atlas.) <https://pmc.ncbi.nlm.nih.gov/articles/PMC13024969/>
     Locator: Abstract, paragraph 1, sentence 1. Retrieved 2026-11-19T01:42:26Z. Source state: Current. Version fingerprint: 8b2e03d1dabfc41a34ca899f60d1dadad278c80683a31ff8d61a00b4df0423f9.
     How it applies: Passage in the source that backs the claim.
  2. Supports: Intrapartum Ultrasound in Vacuum Operative Delivery: A Comprehensive Review and Proposal of the Novel Ultrasound Flexion Point Method (Publication, edition PMC Open Access subset full-text XML from NCBI E-utilities, retrieved 2026-11-19 (2026). An expert commentary and technical proposal. Used: the abstract, the introduction and the instrument-selection discussion. The proposed ultrasound method is not described in the atlas.) <https://pmc.ncbi.nlm.nih.gov/articles/PMC13024969/>
     Locator: 3. Discussion, 3.1. Instrument Selection and Device Considerations, paragraph 1, sentence 3. Retrieved 2026-11-19T01:42:26Z. Source state: Current. Version fingerprint: 8b2e03d1dabfc41a34ca899f60d1dadad278c80683a31ff8d61a00b4df0423f9.
     How it applies: Background statement in the source that cites works the atlas did not read.
  3. Supports: Intrapartum Ultrasound in Vacuum Operative Delivery: A Comprehensive Review and Proposal of the Novel Ultrasound Flexion Point Method (Publication, edition PMC Open Access subset full-text XML from NCBI E-utilities, retrieved 2026-11-19 (2026). An expert commentary and technical proposal. Used: the abstract, the introduction and the instrument-selection discussion. The proposed ultrasound method is not described in the atlas.) <https://pmc.ncbi.nlm.nih.gov/articles/PMC13024969/>
     Locator: 3. Discussion, 3.1. Instrument Selection and Device Considerations, paragraph 1, sentence 2. Retrieved 2026-11-19T01:42:26Z. Source state: Current. Version fingerprint: 8b2e03d1dabfc41a34ca899f60d1dadad278c80683a31ff8d61a00b4df0423f9.
     How it applies: Background statement in the source that cites works the atlas did not read.
  4. Supports: An engineering perspective of vacuum assisted delivery devices in obstetrics: A review (Publication, edition PMC Open Access subset full-text XML from NCBI E-utilities, retrieved 2026-11-19 (2020). An engineering review. Used: the abstract, the design history, the adverse-event passage and the conclusion. Technique detail is not used.) <https://pmc.ncbi.nlm.nih.gov/articles/PMC7780266/>
     Locator: Clinical use of VAD systems, paragraph 6, sentence 4. Retrieved 2026-11-19T01:42:25Z. Source state: Current. Version fingerprint: 3fa8bd408d19470962b39f8526f97760f4c06bf240e0ca82c0b8a642bd7af839.
     How it applies: Background statement in the source that cites works the atlas did not read.
  5. Supports: Intrapartum Ultrasound in Vacuum Operative Delivery: A Comprehensive Review and Proposal of the Novel Ultrasound Flexion Point Method (Publication, edition PMC Open Access subset full-text XML from NCBI E-utilities, retrieved 2026-11-19 (2026). An expert commentary and technical proposal. Used: the abstract, the introduction and the instrument-selection discussion. The proposed ultrasound method is not described in the atlas.) <https://pmc.ncbi.nlm.nih.gov/articles/PMC13024969/>
     Locator: 1. Introduction, paragraph 5, sentence 2. Retrieved 2026-11-19T01:42:26Z. Source state: Current. Version fingerprint: 8b2e03d1dabfc41a34ca899f60d1dadad278c80683a31ff8d61a00b4df0423f9.
     How it applies: Background statement in the source that cites works the atlas did not read.

## Context: Women having an operative vaginal birth with a vacuum extractor, and their babies (engineering review 2020; commentary 2026)

Recipient: Mixed recipients; Population: Women having a vacuum-assisted birth and the babies whose scalp the cup is placed on, as described in an engineering review and an expert commentary. Neither is a study population.; Care setting: Hospital labour wards and delivery rooms, where the reviews place instrumental birth.; Geography: UK and US usage figures are quoted in the sources; not itemised here.; Life stage: Second stage of labour; Gestational age: Not reported; Postnatal age: Not reported; Follow-up: Not applicable (literature review)

### The 2026 commentary reports that historical randomised trials found soft cups to have higher failure and detachment rates than rigid or metal cups, although with fewer scalp injuries. It also reports a comparison in which one cup design had a significantly higher failure rate than conventional vacuum cups (30.1% compared with 19.2%; relative risk 1.58, 95% CI 1.10 to 2.24), and says that randomised and pragmatic trials suggest severe neonatal outcomes remain broadly comparable between disposable and metal devices. The 2020 review adds that unintentional cup detachments are the most prevalent clinical adverse event with these devices.

- Category: Effect
- Evidence stage: Evidence synthesis
- Certainty: Certainty not assessed
- Applies to: Recipient: Mixed recipients; Population: Women having a vacuum-assisted birth and the babies whose scalp the cup is placed on, as described in an engineering review and an expert commentary. Neither is a study population.; Care setting: Hospital labour wards and delivery rooms, where the reviews place instrumental birth.; Geography: UK and US usage figures are quoted in the sources; not itemised here.; Life stage: Second stage of labour; Gestational age: Not reported; Postnatal age: Not reported; Follow-up: Not applicable (literature review)
- Limitations: A second-hand summary of trials the atlas did not read; no cup model is named here. Certainty is not assessed here.
- Last checked: 2026-11-19
- Sources:
  1. Supports: Intrapartum Ultrasound in Vacuum Operative Delivery: A Comprehensive Review and Proposal of the Novel Ultrasound Flexion Point Method (Publication, edition PMC Open Access subset full-text XML from NCBI E-utilities, retrieved 2026-11-19 (2026). An expert commentary and technical proposal. Used: the abstract, the introduction and the instrument-selection discussion. The proposed ultrasound method is not described in the atlas.) <https://pmc.ncbi.nlm.nih.gov/articles/PMC13024969/>
     Locator: 3. Discussion, 3.1. Instrument Selection and Device Considerations, paragraph 2, sentence 2. Retrieved 2026-11-19T01:42:26Z. Source state: Current. Version fingerprint: 8b2e03d1dabfc41a34ca899f60d1dadad278c80683a31ff8d61a00b4df0423f9.
     How it applies: Background statement in the source that cites works the atlas did not read.
  2. Supports: Intrapartum Ultrasound in Vacuum Operative Delivery: A Comprehensive Review and Proposal of the Novel Ultrasound Flexion Point Method (Publication, edition PMC Open Access subset full-text XML from NCBI E-utilities, retrieved 2026-11-19 (2026). An expert commentary and technical proposal. Used: the abstract, the introduction and the instrument-selection discussion. The proposed ultrasound method is not described in the atlas.) <https://pmc.ncbi.nlm.nih.gov/articles/PMC13024969/>
     Locator: 3. Discussion, 3.1. Instrument Selection and Device Considerations, paragraph 2, sentence 4. Retrieved 2026-11-19T01:42:26Z. Source state: Current. Version fingerprint: 8b2e03d1dabfc41a34ca899f60d1dadad278c80683a31ff8d61a00b4df0423f9.
     How it applies: Background statement in the source that cites works the atlas did not read.
  3. Supports: Intrapartum Ultrasound in Vacuum Operative Delivery: A Comprehensive Review and Proposal of the Novel Ultrasound Flexion Point Method (Publication, edition PMC Open Access subset full-text XML from NCBI E-utilities, retrieved 2026-11-19 (2026). An expert commentary and technical proposal. Used: the abstract, the introduction and the instrument-selection discussion. The proposed ultrasound method is not described in the atlas.) <https://pmc.ncbi.nlm.nih.gov/articles/PMC13024969/>
     Locator: 3. Discussion, 3.1. Instrument Selection and Device Considerations, paragraph 2, sentence 6. Retrieved 2026-11-19T01:42:26Z. Source state: Current. Version fingerprint: 8b2e03d1dabfc41a34ca899f60d1dadad278c80683a31ff8d61a00b4df0423f9.
     How it applies: Background statement in the source that cites works the atlas did not read.
  4. Supports: An engineering perspective of vacuum assisted delivery devices in obstetrics: A review (Publication, edition PMC Open Access subset full-text XML from NCBI E-utilities, retrieved 2026-11-19 (2020). An engineering review. Used: the abstract, the design history, the adverse-event passage and the conclusion. Technique detail is not used.) <https://pmc.ncbi.nlm.nih.gov/articles/PMC7780266/>
     Locator: VAD mechanics & performance, paragraph 5, sentence 1. Retrieved 2026-11-19T01:42:25Z. Source state: Current. Version fingerprint: 3fa8bd408d19470962b39f8526f97760f4c06bf240e0ca82c0b8a642bd7af839.
     How it applies: Background statement in the source that cites works the atlas did not read.

### The 2020 review traces how the devices changed: pliable cups made from elastomeric materials were introduced in the 1970s, driven by concern that rigid metal cups could lead to scalp trauma on the infant, and single-use instruments followed in the 1990s, catalysed by moves to reduce transmissible infection. It adds that much of this evidence is circumstantial.

- Category: Description
- Evidence stage: Evidence synthesis
- Certainty: Certainty not assessed
- Applies to: Recipient: Mixed recipients; Population: Women having a vacuum-assisted birth and the babies whose scalp the cup is placed on, as described in an engineering review and an expert commentary. Neither is a study population.; Care setting: Hospital labour wards and delivery rooms, where the reviews place instrumental birth.; Geography: UK and US usage figures are quoted in the sources; not itemised here.; Life stage: Second stage of labour; Gestational age: Not reported; Postnatal age: Not reported; Follow-up: Not applicable (literature review)
- Limitations: A historical account by the review authors; the claim about circumstantial evidence is theirs and is not tested here.
- Last checked: 2026-11-19
- Sources:
  1. Supports: An engineering perspective of vacuum assisted delivery devices in obstetrics: A review (Publication, edition PMC Open Access subset full-text XML from NCBI E-utilities, retrieved 2026-11-19 (2020). An engineering review. Used: the abstract, the design history, the adverse-event passage and the conclusion. Technique detail is not used.) <https://pmc.ncbi.nlm.nih.gov/articles/PMC7780266/>
     Locator: Design evolution of VAD devices, paragraph 4, sentence 1. Retrieved 2026-11-19T01:42:25Z. Source state: Current. Version fingerprint: 3fa8bd408d19470962b39f8526f97760f4c06bf240e0ca82c0b8a642bd7af839.
     How it applies: Passage in the source that backs the claim.
  2. Supports: An engineering perspective of vacuum assisted delivery devices in obstetrics: A review (Publication, edition PMC Open Access subset full-text XML from NCBI E-utilities, retrieved 2026-11-19 (2020). An engineering review. Used: the abstract, the design history, the adverse-event passage and the conclusion. Technique detail is not used.) <https://pmc.ncbi.nlm.nih.gov/articles/PMC7780266/>
     Locator: Design evolution of VAD devices, paragraph 5, sentence 1. Retrieved 2026-11-19T01:42:25Z. Source state: Current. Version fingerprint: 3fa8bd408d19470962b39f8526f97760f4c06bf240e0ca82c0b8a642bd7af839.
     How it applies: Passage in the source that backs the claim.
  3. Supports: An engineering perspective of vacuum assisted delivery devices in obstetrics: A review (Publication, edition PMC Open Access subset full-text XML from NCBI E-utilities, retrieved 2026-11-19 (2020). An engineering review. Used: the abstract, the design history, the adverse-event passage and the conclusion. Technique detail is not used.) <https://pmc.ncbi.nlm.nih.gov/articles/PMC7780266/>
     Locator: Design evolution of VAD devices, paragraph 6, sentence 3. Retrieved 2026-11-19T01:42:25Z. Source state: Current. Version fingerprint: 3fa8bd408d19470962b39f8526f97760f4c06bf240e0ca82c0b8a642bd7af839.
     How it applies: Passage in the source that backs the claim.

### A 2020 engineering review describes a vacuum assisted delivery device as a suction cup placed on the baby's scalp, with negative pressure applied by a manual or electric pump. It reports that, after more than 50 years in clinical practice, these devices remain a mainstay in physically assisting birth and are sometimes preferred over forceps for ease of use and reduced maternal morbidity.

- Category: Description
- Evidence stage: Evidence synthesis
- Certainty: Certainty not assessed
- Applies to: Recipient: Mixed recipients; Population: Women having a vacuum-assisted birth and the babies whose scalp the cup is placed on, as described in an engineering review and an expert commentary. Neither is a study population.; Care setting: Hospital labour wards and delivery rooms, where the reviews place instrumental birth.; Geography: UK and US usage figures are quoted in the sources; not itemised here.; Life stage: Second stage of labour; Gestational age: Not reported; Postnatal age: Not reported; Follow-up: Not applicable (literature review)
- Limitations: General statements in a review that cites works the atlas did not read. The atlas gives no indication, technique or advice.
- Last checked: 2026-11-19
- Sources:
  1. Supports: An engineering perspective of vacuum assisted delivery devices in obstetrics: A review (Publication, edition PMC Open Access subset full-text XML from NCBI E-utilities, retrieved 2026-11-19 (2020). An engineering review. Used: the abstract, the design history, the adverse-event passage and the conclusion. Technique detail is not used.) <https://pmc.ncbi.nlm.nih.gov/articles/PMC7780266/>
     Locator: Introduction, paragraph 8, sentence 2. Retrieved 2026-11-19T01:42:25Z. Source state: Current. Version fingerprint: 3fa8bd408d19470962b39f8526f97760f4c06bf240e0ca82c0b8a642bd7af839.
     How it applies: Passage in the source that backs the claim.
  2. Supports: An engineering perspective of vacuum assisted delivery devices in obstetrics: A review (Publication, edition PMC Open Access subset full-text XML from NCBI E-utilities, retrieved 2026-11-19 (2020). An engineering review. Used: the abstract, the design history, the adverse-event passage and the conclusion. Technique detail is not used.) <https://pmc.ncbi.nlm.nih.gov/articles/PMC7780266/>
     Locator: Abstract, paragraph 1, sentence 2. Retrieved 2026-11-19T01:42:25Z. Source state: Current. Version fingerprint: 3fa8bd408d19470962b39f8526f97760f4c06bf240e0ca82c0b8a642bd7af839.
     How it applies: Passage in the source that backs the claim.

### The 2020 review says the technical literature on vacuum assisted delivery devices is scarce, disparate and often difficult to source, that there is minimal evidence to inform device design or clinical use, and that there is an urgent need to evaluate the performance of these devices. The 2026 commentary reports a review of randomised trials in which sonographic knowledge of fetal head position before operative vaginal birth did not appear to reduce the incidence of failed procedures. It also proposes an ultrasound-based method and says that proposal requires prospective validation through randomised controlled trials. The atlas does not describe that method.

- Category: Description
- Evidence stage: Evidence synthesis
- Certainty: Certainty not assessed
- Applies to: Recipient: Mixed recipients; Population: Women having a vacuum-assisted birth and the babies whose scalp the cup is placed on, as described in an engineering review and an expert commentary. Neither is a study population.; Care setting: Hospital labour wards and delivery rooms, where the reviews place instrumental birth.; Geography: UK and US usage figures are quoted in the sources; not itemised here.; Life stage: Second stage of labour; Gestational age: Not reported; Postnatal age: Not reported; Follow-up: Not applicable (literature review)
- Limitations: Statements by the review authors about the state of the evidence. The commentary's method is a proposal, not a tested practice. Certainty is not assessed here.
- Last checked: 2026-11-19
- Sources:
  1. Supports: An engineering perspective of vacuum assisted delivery devices in obstetrics: A review (Publication, edition PMC Open Access subset full-text XML from NCBI E-utilities, retrieved 2026-11-19 (2020). An engineering review. Used: the abstract, the design history, the adverse-event passage and the conclusion. Technique detail is not used.) <https://pmc.ncbi.nlm.nih.gov/articles/PMC7780266/>
     Locator: Introduction, paragraph 9, sentence 3. Retrieved 2026-11-19T01:42:25Z. Source state: Current. Version fingerprint: 3fa8bd408d19470962b39f8526f97760f4c06bf240e0ca82c0b8a642bd7af839.
     How it applies: Passage in the source that backs the claim.
  2. Supports: An engineering perspective of vacuum assisted delivery devices in obstetrics: A review (Publication, edition PMC Open Access subset full-text XML from NCBI E-utilities, retrieved 2026-11-19 (2020). An engineering review. Used: the abstract, the design history, the adverse-event passage and the conclusion. Technique detail is not used.) <https://pmc.ncbi.nlm.nih.gov/articles/PMC7780266/>
     Locator: Conclusion, paragraph 2, sentence 3. Retrieved 2026-11-19T01:42:25Z. Source state: Current. Version fingerprint: 3fa8bd408d19470962b39f8526f97760f4c06bf240e0ca82c0b8a642bd7af839.
     How it applies: Passage in the source that backs the claim.
  3. Supports: Intrapartum Ultrasound in Vacuum Operative Delivery: A Comprehensive Review and Proposal of the Novel Ultrasound Flexion Point Method (Publication, edition PMC Open Access subset full-text XML from NCBI E-utilities, retrieved 2026-11-19 (2026). An expert commentary and technical proposal. Used: the abstract, the introduction and the instrument-selection discussion. The proposed ultrasound method is not described in the atlas.) <https://pmc.ncbi.nlm.nih.gov/articles/PMC13024969/>
     Locator: 1. Introduction, paragraph 9, sentence 2. Retrieved 2026-11-19T01:42:26Z. Source state: Current. Version fingerprint: 8b2e03d1dabfc41a34ca899f60d1dadad278c80683a31ff8d61a00b4df0423f9.
     How it applies: Background statement in the source that cites works the atlas did not read.
  4. Supports: Intrapartum Ultrasound in Vacuum Operative Delivery: A Comprehensive Review and Proposal of the Novel Ultrasound Flexion Point Method (Publication, edition PMC Open Access subset full-text XML from NCBI E-utilities, retrieved 2026-11-19 (2026). An expert commentary and technical proposal. Used: the abstract, the introduction and the instrument-selection discussion. The proposed ultrasound method is not described in the atlas.) <https://pmc.ncbi.nlm.nih.gov/articles/PMC13024969/>
     Locator: Abstract, paragraph 1, sentence 9. Retrieved 2026-11-19T01:42:26Z. Source state: Current. Version fingerprint: 8b2e03d1dabfc41a34ca899f60d1dadad278c80683a31ff8d61a00b4df0423f9.
     How it applies: Passage in the source that backs the claim.

## Context: US federal device rules for fetal vacuum extractors, not a study population

Recipient: No clinical recipient; Population: No population. The text identifies and classifies a generic type of medical device in United States rules.; Care setting: Not applicable.; Geography: United States.; Life stage: Not applicable; Follow-up: Not applicable (regulatory context).

### Title 21 of the Code of Federal Regulations, section 884.4340 (eCFR text returned for 1 November 2026), identifies a fetal vacuum extractor as a device used to facilitate delivery, applying traction to the fetal head in the birth canal by a suction cup attached to the scalp and powered by an external vacuum source, and classifies the generic type as Class II (performance standards). The openFDA classification record for the device lists the product code HDB, the name Extractor, Vacuum, Fetal, and device class 2.

- Category: Regulatory
- Evidence stage: Evidence stage not applicable
- Certainty: Certainty not assessed
- Applies to: Recipient: No clinical recipient; Population: No population. The text identifies and classifies a generic type of medical device in United States rules.; Care setting: Not applicable.; Geography: United States.; Life stage: Not applicable; Follow-up: Not applicable (regulatory context).
- Limitations: A federal classification of a generic device type, not a statement about any brand or about safety or benefit in use. openFDA states that its data are unvalidated and not for medical decisions. The record lists many 510(k) numbers; none was read. The eCFR is an editorial compilation, not the legal edition.
- Last checked: 2026-11-19
- Sources:
  1. Supports: 21 CFR 884.4340 Fetal vacuum extractor (eCFR) (Regulator, edition eCFR section text as returned by the versioner API for 2026-11-01. The first request, made without a gzip encoding header, was not accepted; the committed file is the decompressed response to the second request.) <https://www.ecfr.gov/api/versioner/v1/full/2026-11-01/title-21.xml?part=884&section=884.4340>
     Locator: 884.4340 (a) Identification, sentence 1. Retrieved 2026-11-19T01:52:55Z. Source state: Current. Version fingerprint: c3eec01f4ee3d562b635c2ec7e10b2799eea5be930268eb96a6123ea87d5b24d.
     How it applies: Passage in the source that backs the claim.
  2. Supports: 21 CFR 884.4340 Fetal vacuum extractor (eCFR) (Regulator, edition eCFR section text as returned by the versioner API for 2026-11-01. The first request, made without a gzip encoding header, was not accepted; the committed file is the decompressed response to the second request.) <https://www.ecfr.gov/api/versioner/v1/full/2026-11-01/title-21.xml?part=884&section=884.4340>
     Locator: 884.4340 (a) Identification, sentence 2, part 1. Retrieved 2026-11-19T01:52:55Z. Source state: Current. Version fingerprint: c3eec01f4ee3d562b635c2ec7e10b2799eea5be930268eb96a6123ea87d5b24d.
     How it applies: Passage in the source that backs the claim.
  3. Supports: 21 CFR 884.4340 Fetal vacuum extractor (eCFR) (Regulator, edition eCFR section text as returned by the versioner API for 2026-11-01. The first request, made without a gzip encoding header, was not accepted; the committed file is the decompressed response to the second request.) <https://www.ecfr.gov/api/versioner/v1/full/2026-11-01/title-21.xml?part=884&section=884.4340>
     Locator: 884.4340 (a) Identification, sentence 2, part 2. Retrieved 2026-11-19T01:52:55Z. Source state: Current. Version fingerprint: c3eec01f4ee3d562b635c2ec7e10b2799eea5be930268eb96a6123ea87d5b24d.
     How it applies: Passage in the source that backs the claim.
  4. Supports: 21 CFR 884.4340 Fetal vacuum extractor (eCFR) (Regulator, edition eCFR section text as returned by the versioner API for 2026-11-01. The first request, made without a gzip encoding header, was not accepted; the committed file is the decompressed response to the second request.) <https://www.ecfr.gov/api/versioner/v1/full/2026-11-01/title-21.xml?part=884&section=884.4340>
     Locator: 884.4340 (b) Classification. Retrieved 2026-11-19T01:52:55Z. Source state: Current. Version fingerprint: c3eec01f4ee3d562b635c2ec7e10b2799eea5be930268eb96a6123ea87d5b24d.
     How it applies: Passage in the source that backs the claim.
  5. Supports: openFDA device classification records matching the device name vacuum extractor (product code HDB, Extractor, Vacuum, Fetal) (Dataset, edition openFDA device classification search result fetched on 2026-11-19 (openFDA data last updated 2026-11-09). It holds one record: Extractor, Vacuum, Fetal (HDB).) <https://api.fda.gov/device/classification.json?search=device_name:%22vacuum%22+AND+device_name:%22extractor%22&limit=10>
     Locator: results\[0\].device\_name. Retrieved 2026-11-19T01:52:34Z. Source state: Current. Version fingerprint: 471215c51af1b0dce8833165dedaf40a89a49e19f5ae3fbf5ff38acbc0c2b07d.
     How it applies: Passage in the source that backs the claim.
  6. Supports: openFDA device classification records matching the device name vacuum extractor (product code HDB, Extractor, Vacuum, Fetal) (Dataset, edition openFDA device classification search result fetched on 2026-11-19 (openFDA data last updated 2026-11-09). It holds one record: Extractor, Vacuum, Fetal (HDB).) <https://api.fda.gov/device/classification.json?search=device_name:%22vacuum%22+AND+device_name:%22extractor%22&limit=10>
     Locator: results\[0\].product\_code. Retrieved 2026-11-19T01:52:34Z. Source state: Current. Version fingerprint: 471215c51af1b0dce8833165dedaf40a89a49e19f5ae3fbf5ff38acbc0c2b07d.
     How it applies: Passage in the source that backs the claim.
  7. Supports: openFDA device classification records matching the device name vacuum extractor (product code HDB, Extractor, Vacuum, Fetal) (Dataset, edition openFDA device classification search result fetched on 2026-11-19 (openFDA data last updated 2026-11-09). It holds one record: Extractor, Vacuum, Fetal (HDB).) <https://api.fda.gov/device/classification.json?search=device_name:%22vacuum%22+AND+device_name:%22extractor%22&limit=10>
     Locator: results\[0\].device\_class. Retrieved 2026-11-19T01:52:34Z. Source state: Current. Version fingerprint: 471215c51af1b0dce8833165dedaf40a89a49e19f5ae3fbf5ff38acbc0c2b07d.
     How it applies: Passage in the source that backs the claim.
  8. Supports: openFDA device classification records matching the device name vacuum extractor (product code HDB, Extractor, Vacuum, Fetal) (Dataset, edition openFDA device classification search result fetched on 2026-11-19 (openFDA data last updated 2026-11-09). It holds one record: Extractor, Vacuum, Fetal (HDB).) <https://api.fda.gov/device/classification.json?search=device_name:%22vacuum%22+AND+device_name:%22extractor%22&limit=10>
     Locator: results\[0\].regulation\_number. Retrieved 2026-11-19T01:52:34Z. Source state: Current. Version fingerprint: 471215c51af1b0dce8833165dedaf40a89a49e19f5ae3fbf5ff38acbc0c2b07d.
     How it applies: Passage in the source that backs the claim.

## Context: Women having an operative vaginal birth with a vacuum extractor or forceps (2026 commentary)

Recipient: Maternal; Population: Women having an instrumental vaginal birth; the commentary summarises Cochrane review findings on perineal trauma and success of vaginal birth for vacuum extraction compared with forceps.; Care setting: Hospital labour wards.; Geography: Not stated in the source; Life stage: Second stage of labour; Gestational age: Not reported; Postnatal age: Not reported; Follow-up: Not applicable (literature commentary)

### A 2026 expert commentary says that vacuum extraction has gained popularity because of a lower risk of maternal perineal trauma than with forceps, and, citing randomised evidence summarised by Cochrane reviews, that forceps are more likely to achieve vaginal birth than vacuum extraction but carry higher rates of severe maternal perineal trauma.

- Category: Effect
- Evidence stage: Evidence synthesis
- Certainty: Certainty not assessed
- Applies to: Recipient: Maternal; Population: Women having an instrumental vaginal birth; the commentary summarises Cochrane review findings on perineal trauma and success of vaginal birth for vacuum extraction compared with forceps.; Care setting: Hospital labour wards.; Geography: Not stated in the source; Life stage: Second stage of labour; Gestational age: Not reported; Postnatal age: Not reported; Follow-up: Not applicable (literature commentary)
- Limitations: The commentary summarises reviews and trials it does not reproduce; the atlas did not read them. No figures are given in the passage. Certainty is not assessed here.
- Last checked: 2026-11-19
- Sources:
  1. Supports: Intrapartum Ultrasound in Vacuum Operative Delivery: A Comprehensive Review and Proposal of the Novel Ultrasound Flexion Point Method (Publication, edition PMC Open Access subset full-text XML from NCBI E-utilities, retrieved 2026-11-19 (2026). An expert commentary and technical proposal. Used: the abstract, the introduction and the instrument-selection discussion. The proposed ultrasound method is not described in the atlas.) <https://pmc.ncbi.nlm.nih.gov/articles/PMC13024969/>
     Locator: 1. Introduction, paragraph 1, sentence 5. Retrieved 2026-11-19T01:42:26Z. Source state: Current. Version fingerprint: 8b2e03d1dabfc41a34ca899f60d1dadad278c80683a31ff8d61a00b4df0423f9.
     How it applies: Passage in the source that backs the claim.
  2. Supports: Intrapartum Ultrasound in Vacuum Operative Delivery: A Comprehensive Review and Proposal of the Novel Ultrasound Flexion Point Method (Publication, edition PMC Open Access subset full-text XML from NCBI E-utilities, retrieved 2026-11-19 (2026). An expert commentary and technical proposal. Used: the abstract, the introduction and the instrument-selection discussion. The proposed ultrasound method is not described in the atlas.) <https://pmc.ncbi.nlm.nih.gov/articles/PMC13024969/>
     Locator: 3. Discussion, 3.1. Instrument Selection and Device Considerations, paragraph 1, sentence 2. Retrieved 2026-11-19T01:42:26Z. Source state: Current. Version fingerprint: 8b2e03d1dabfc41a34ca899f60d1dadad278c80683a31ff8d61a00b4df0423f9.
     How it applies: Background statement in the source that cites works the atlas did not read.

## Connected pages

- The birth procedure in which the device is used: Operative vaginal birth (forceps and vacuum) (`maternal:operative-vaginal-birth`)
- Scalp injuries are among the reported risks to the baby: Birth injuries to the baby (neonatal birth trauma) (`maternal:birth-injuries`)
- Maternal perineal trauma is compared between vacuum and forceps in the 2026 commentary: Perineal tear and obstetric anal sphincter injury (`maternal:perineal-tear-and-obstetric-anal-sphincter-injury`)
- Filed under this front: Procedures and birth care (`maternal:front-procedures-and-birth-care`)

## Provenance

Content hash: 2349312166dfea471a882195f56bf07f38774d2b465e84977561c31d23d54eb2
Last substantive edit: 2026-11-19
Evidence last checked: 2026-11-19